Article
Dual-source cardiac computed tomography: image quality and dose considerations.
Department of Cardiology, University Erlangen-Nuernberg, Ulmenweg 18, 91054 Erlangen, Germany.
European Radiology (impact factor:
3.22).
07/2008;
18(6):1188-98.
DOI:10.1007/s00330-008-0883-3
Source: PubMed
-
Article: ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart.
[show abstract] [hide abstract]
ABSTRACT: Subsecond spiral computed tomography (CT) offers great potential for improving heart imaging. The new multi-row detector technology adds significantly to this potential. We therefore developed and validated dedicated cardiac reconstruction algorithms for imaging the heart with subsecond multi-slice spiral CT utilizing electrocardiogram (ECG) information. The single-slice cardiac z-interpolation algorithms 180 degrees CI and 180 degrees CD [Med. Phys. 25, 2417-2431 (1998)] were generalized to allow imaging of the heart for M-slice scanners. Two classes of algorithms were investigated: 180 degrees MCD (multi-slice cardio delta), a partial scan reconstruction of 180 degrees + delta data with a < phi (fan angle) resulting in effective scan times of 250 ms (central ray) when a 0.5 s rotation mode is available, and 180 degrees MCI (multi-slice cardio interpolation), a piecewise weighted interpolation between successive spiral data segments belonging to the same heart phase, potentially providing a relative temporal resolution of 12.5% of the heart cycle when a four-slice scanner is used and the table increment is chosen to be greater than or equal to the collimated slice thickness. Data segments are selected by correlation with the simultaneously recorded ECG signal. Theoretical studies, computer simulations, as well as patient measurements were carried out for a multi-slice scanner providing M = 4 slices to evaluate these new approaches and determine the optimal scan protocol. Both algorithms, 180 degrees MCD and 180 degrees MCI, provide significant improvements in image quality, including extremely arythmic cases. Artifacts in the reconstructed images as well as in 3D displays such as multiplanar reformations were largely reduced as compared to the standard z-interpolation algorithm 180 degrees MLI (multi-slice linear interpolation). Image quality appears adequate for precise calcium scoring and CT angiography of the coronary arteries with conventional subsecond multislice spiral CT. It turned out that for heart rates fH > or = 70 min(-1) the partial scan approach 180 degrees MCD yields unsatisfactory results as compared to 180 degrees MCI. Our theoretical considerations show that a freely selectable scanner rotation time chosen as a function of the patient's heart rate, would further improve the relative temporal resolution and thus further reduce motion artifacts. In our case an additional 0.6 s mode besides the available 0.5 s mode would be very helpful. Moreover, if technically feasible, lower rotation times such as 0.3 s or even less would result in improved image quality. The use of multi-slice techniques for cardiac CT together with the new z-interpolation methods improves the quality of heart imaging significantly. The high temporal resolution of 180 degrees MCI is adequate for spatial and temporal tracking of anatomic structures of the heart (4D reconstruction).Medical Physics 08/2000; 27(8):1881-902. · 2.83 Impact Factor -
Article: Computed tomography coronary angiography.
[show abstract] [hide abstract]
ABSTRACT: Recent developments in computed tomography technology have made imaging of the coronary arteries possible. All the same, the rapid motion and small dimensions of the coronary vessels make coronary computed tomography angiography (coronary CTA) challenging. With the last generations of 16- and 64-slice computed tomography and adequate patient preparation (which includes lowering of the heart rate), rates of sensitivity ranging from 83% to 99% and specificity between 93% and 98% have been reported for the detection of coronary artery stenoses in comparison with invasive coronary angiography. The high negative predictive value (95% to 100%) found in these studies suggests that coronary CTA may be a useful diagnostic technique to rule out the presence of coronary stenoses in selected patients, especially those with a rather low pretest likelihood of disease. Imaging of coronary artery bypass grafts is reliable, but clinical applications can be hampered by difficulties in assessing the native coronary arteries in patients after undergoing bypass because of their often-severe calcification. The detection of in-stent restenosis is made difficult by artifacts caused by metal, especially in smaller stents. Finally, initial reports that coronary CTA allows the detection and, to a certain extent, also the characterization and quantification of noncalcified coronary arteriosclerotic plaque are interesting, but they currently do not provide sufficient data to support clinical applications in the context of risk stratification.Journal of the American College of Cardiology 12/2006; 48(10):1919-28. · 14.16 Impact Factor -
Article: Diagnostic performance of multidetector CT angiography for assessment of coronary artery disease: meta-analysis.
[show abstract] [hide abstract]
ABSTRACT: To review the literature on the diagnostic performance of multidetector computed tomographic (CT) angiography for assessment of symptomatic coronary artery disease, with conventional coronary angiography as the reference standard. A PubMed and manual search of the literature published between January 1998 and May 2006 on use of multidetector CT angiography compared with coronary angiography in patients with symptomatic coronary artery disease was performed. Summary estimates of diagnostic odds ratio, sensitivity, and specificity were calculated. Random-effects models were used to compare the diagnostic performance of four-, 16-, and 64-detector CT angiographic units, and the proportion of nonassessable coronary arterial segments was evaluated. Fifty-four studies were included in the meta-analysis: 22 studies with four-detector CT angiography, 26 with 16-detector CT angiography, and six with 64-detector CT angiography. The pooled sensitivity and specificity for detecting a greater than 50% stenosis per segment were 0.93 (95% confidence interval [CI]: 0.88, 0.97) and 0.96 (95% CI: 0.96, 0.97) for 64-detector CT angiography, 0.83 (95% CI: 0.76, 0.90) and 0.96 (95% CI: 0.95, 0.97) for 16-detector CT angiography, and 0.84 (95% CI: 0.81, 0.88) and 0.93 (95% CI: 0.91, 0.95) for four-detector CT angiography, respectively. Results of regression analysis indicated that the diagnostic performance significantly improved with the newer generations of multidetector CT scanners (64- and 16-detector vs four-detector units), adjusted for exclusion of nonassessable segments, and contrast agent concentration used (P < .05). Simultaneously, the nonassessable proportion of segments significantly decreased with the newer generations of multidetector CT scanners, adjusted for heart rate, prevalence of significant disease, and mean age. With the newer generations of multidetector CT scanners, the diagnostic performance for the assessment of coronary artery disease has significantly improved, and the proportion of nonassessable segments has decreased.Radiology 09/2007; 244(2):419-28. · 5.73 Impact Factor
Data provided are for informational purposes only. Although carefully collected, accuracy cannot be guaranteed.
The impact factor represents a rough estimation of the journal's impact factor and does not reflect the actual
current impact factor.
Publisher conditions are provided by RoMEO. Differing provisions from the publisher's actual policy or licence
agreement may be applicable.
Keywords
cardiac CT
cardiac imaging
clinical decision-making
Computed tomography
coronary arteries
detecting coronary artery stenoses
diagnostic image quality
Dose issues
dual-source CT
effective scan time
heart rate
image quality
large trials
limited temporal resolution
lower radiation exposure
prominently coronary CT angiography
routine beta blockade
temporal resolution
three-dimensional dose distributions
ventricular function