Intracardiac QT integral on far-field ICD electrogram predicts sustained ventricular tachyarrhythmias in ICD patients

Division of Cardiology, Department of Medicine, Johns Hopkins Hospital, Baltimore, Maryland, USA.
Heart rhythm: the official journal of the Heart Rhythm Society (Impact Factor: 5.08). 07/2011; 8(12):1889-94. DOI: 10.1016/j.hrthm.2011.07.029
Source: PubMed


Prediction of sustained ventricular tachycardia (VT)/ventricular fibrillation (VF) could help to guide preventive interventions in at-risk patients. The QRST integral (∫QT) reflects intrinsic repolarization properties.
The objective of this study was to determine whether intracardiac ∫QT predicts VT/VF in the next few months in patients with implantable cardioverter defibrillators (ICDs).
Far-field (FF) and near-field (NF) right ventricular intracardiac electrograms (EGMs) were recorded via telemetry in 46 patients with structural heart disease and ICDs implanted for secondary prevention of sudden cardiac death. Epochs of 4.9 ± 0.4 minutes during sinus rhythm (mean heart rate 70.9 ± 15.2 beats/min) and ventricular pacing at 105 beats/min were analyzed. Mean ∫QT was calculated on FF and NF EGMs as the algebraic sum of areas under the QRST curve and adjusted by mean heart rate. Patients were followed up for at least 3 months. True VT/VF events treated by the ICD served as the end point.
During a mean follow-up of 4.6 months, 22 patients (48%) were treated for VT/VF. Unadjusted and adjusted by heart rate, FF EGM ∫QT in sinus rhythm was a significant predictor of VT/VF (unadjusted ∫QT hazard ratio 1.007; 95% confidence interval 1.002 to 1.0013; P = .007; adjusted ∫QT hazard ratio 1.68; 95% confidence interval 1.19 to 2.36; P = .002). The highest quartile of intracardiac ∫QT predicted VT/VF (log-rank test P = .042) and identified patients at risk with a specificity of 86% and positive predictive value of 73%.
Increased intracardiac FF EGM ∫QT predicts VT/VF in patients with structural heart disease and secondary prevention ICDs.

9 Reads
  • [Show abstract] [Hide abstract]
    ABSTRACT: We constructed an intracardiac vectorcardiogram from 3 configurations of intracardiac cardiovertor defibrilator (ICD) electrograms (EGMs). Six distinctive 3 lead combinations were selected out of five leads: can to right ventricular coil (RVC); RVC to superior vena cava coil (SVC); atrial lead tip (A-tip) to right ventricular (RV)-ring; can to RV-ring; RV-tip to RVC, in a patient with dual chamber ICD. Surface spatial QRS-T angle (119.8°) was similar to intracardiac spatial QRS-T angle derived from ICD EGMs combination A (101.3°), B (96.1°), C (92.8°), D (95.2), E (99.0), F (96.2) and median (101.5). Future validation of the novel method is needed. Copyright © 2015. Published by Elsevier Inc.
    Journal of electrocardiology 05/2015; DOI:10.1016/j.jelectrocard.2015.05.008 · 1.36 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: The purposes of this study were to characterize and quantify concordance between consecutive atrial and ventricular activation time points through analysis of phases and to explore its association with outcomes in patients with implantable cardioverter-defibrillator (ICD). Patients with structural heart disease and dual-chamber ICDs underwent 5min baseline right ventricular (V) near-field and atrial (A) electrogram (EGM) recording. The cross-dependencies of phase dynamics of the changes in consecutive A (AA') and V (VV') were quantified and the AV phase dependency index was determined. In Cox regression analysis, a high AV phase index (in the highest quartile, >0.259) was significantly associated with higher risk of ventricular tachyarrhythmias (HR 2.84; 95% CI 1.05-7.67; P=0.04). In conclusion, in ICD patients with structural heart disease, high sinus AV phase dependency index on EGM is associated with the risk of ventricular arrhythmia. Copyright © 2015. Published by Elsevier Inc.
    Journal of electrocardiology 08/2015; DOI:10.1016/j.jelectrocard.2015.08.018 · 1.36 Impact Factor