Article

Evaluation of Respiratory Motion Effect on Defect Detection in Myocardial Perfusion SPECT: A Simulation Study.

Department of Biomedical Imaging and Radiological Sciences, National Yang-Ming University, Taipei, Taiwan.
IEEE Transactions on Nuclear Science (Impact Factor: 1.46). 06/2009; 56(3):671-676. DOI: 10.1109/TNS.2009.2015446
Source: PubMed

ABSTRACT The objective of this study is to investigate the effects of respiratory motion (RM) on defect detection in Tc-99m sestamibi myocardial perfusion SPECT (MPS) using a phantom population that includes patient variability. Three RM patterns are included, namely breath-hold, slightly enhanced normal breathing, and deep breathing. For each RM pattern, six 4-D NCAT phantoms were generated, each with anatomical variations. Anterior, lateral and inferior myocardial defects with different sizes and contrasts were inserted. Noise-free SPECT projections were simulated using an analytical projector. Poisson noise was then added to generate noisy realizations. The projection data were reconstructed using the OS-EM algorithm with 1 and 4 subsets/iteration and at 1, 2, 3, 5, 7, and 10 iterations. Short-axis images centered at the centroid of the myocardial defect were extracted, and the channelized Hotelling observer (CHO) was applied for the detection of the defect. The CHO results show that the value of the area under the receiver operating characteristics (ROC) curve (AUC) is affected by the RM amplitude. For all the defect sizes and contrasts studied, the highest or optimal AUC values indicate maximum detectability decrease with the increase of the RM amplitude. With no respiration, the ranking of the optimal AUC value in decreasing order is anterior then lateral, and finally inferior defects. The AUC value of the lateral defect drops more severely as the RM amplitude increases compared to other defect locations. Furthermore, as the RM amplitude increases, the AUC values of the smaller defects drop more quickly than the larger ones. We demonstrated that RM affects defect detectability of MPS imaging. The results indicate that developments of optimal data acquisition methods and RM correction methods are needed to improve the defect detectability in MPS.

Download full-text

Full-text

Available from: Jyh-Cheng Chen, Jun 30, 2015
0 Followers
 · 
109 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: The complex interfaces created by migration not only challenge core beliefs about the purpose of career guidance and counseling but also about the precise nature and level of the support required for migrants. However, the issue has had little academic attention. While traditional theories informing the practice of career guidance and counseling implicitly reflect a commitment to social equality, they were developed in relatively homogenous Western capitalist contexts that were strongly individualized, masculine, secular, action and future focused. As career guidance and counseling services in more individualized countries seek to meet the needs of a growing number of clients from more collectivist cultures, the universal relevance of such models is increasingly open to challenge. This article explores three core challenges posed by migration for the theory and practice of career guidance and counseling. The first relates to the very purpose of these types of services for migrants; the second, to the way in which migration requires a realignment of the relationship of theory with practice; and the third relates to the need to redefine service delivery models for this client group. The extent to which notions of “quality” in services are culturally dependent will also be considered, together with the potential for general service improvements being stimulated by the provision of high quality services for migrants.
    Journal of Vocational Behavior 06/2011; 78(3):325-333. DOI:10.1016/j.jvb.2011.03.012 · 2.03 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: To better understand fundamental issues, perception studies of the fusion display would best be performed with a panel of lesions of variable location, size, intensity, and background. There are compelling reasons to use synthetic images that contain artificial lesions for perception research. A consideration of how to obtain this panel of lesions is the nucleus of the present review. This article is a conjoint effort of 3 groups that have joined together to review results from work that they and others have performed. The techniques we review include (1) substitution of lesions into a preexisting image matrix (either using actual prior patient-derived lesions or mathematically modeled artificial lesions), (2) addition of images (either in the attenuation-corrected image space or at an earlier stage before image reconstruction), and (3) simulation of the entire patient image. A judicious combination of the techniques discussed in this review may represent the most efficient pathway of simulating statistically varied but realistic appearing lesions.
    Seminars in nuclear medicine 11/2011; 41(6):437-48. DOI:10.1053/j.semnuclmed.2011.06.007 · 3.13 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Respiratory and cardiac motions can degrade myocardial perfusion SPECT (MPS) image quality and reduce defect detection and quantitative accuracy. In this study, we developed a dual respiratory and cardiac gating system for a high-resolution fully stationary cardiac SPECT scanner in order to improve the image quality and defect detection. Respiratory motion was monitored using a compressive sensor pillow connected to a dual respiratory–cardiac gating box, which sends cardiac triggers only during end-expiration phases to the single cardiac trigger input on the SPECT scanners. The listmode data were rebinned retrospectively into end-expiration frames for respiratory motion reduction or eight cardiac gates only during end-expiration phases to compensate for both respiratory and cardiac motions. The proposed method was first validated on a motion phantom in the presence and absence of multiple perfusion defects, and then applied on 11 patient studies with and without perfusion defects. In the normal phantom studies, the end-expiration gated SPECT (EXG-SPECT) reduced respiratory motion blur and increased myocardium to blood pool contrast by 51.2% as compared to the ungated images. The proposed method also yielded an average of 11.2% increase in myocardium to defect contrast as compared to the ungated images in the phantom studies with perfusion defects. In the patient studies, EXG-SPECT significantly improved the myocardium to blood pool contrast (p < 0.005) by 24% on average as compared to the ungated images, and led to improved perfusion uniformity across segments on polar maps for normal patients. For a patient with defect, EXG-SPECT improved the defect contrast and definition. The dual respiratory–cardiac gating further reduced the blurring effect, increased the myocardium to blood pool contrast significantly by 36% (p < 0.05) compared to EXG-SPECT, and further improved defect characteristics and visualization of fine structures at the expense of increased noise on the patient with defect. The results showed that the proposed methods can effectively reduce motion blur in the images caused by both respiratory and cardiac motions, which may lead to more accurate defect detection and quantifications. This approach can be easily adapted in routine clinical practice on currently available commercial systems.
    Physics in Medicine and Biology 09/2014; 59(20):6267-6287. DOI:10.1088/0031-9155/59/20/6267 · 2.92 Impact Factor