Intratissue Refractive Index Shaping (IRIS) of the Cornea and Lens Using a Low-Pulse-Energy Femtosecond Laser Oscillator

Institute of Optics, University of Rochester, Rochester, New York, USA.
Investigative ophthalmology & visual science (Impact Factor: 3.66). 12/2008; 49(12):5332-9. DOI: 10.1167/iovs.08-1921
Source: PubMed

ABSTRACT To assess the optical effect of high-repetition-rate, low-energy femtosecond laser pulses on lightly fixed corneas and lenses.
Eight corneas and eight lenses were extracted postmortem from normal, adult cats. They were lightly fixed and stored in a solution that minimized swelling and opacification. An 800-nm Ti:Sapphire femtosecond laser oscillator with a 27-fs pulse duration and 93-MHz repetition rate was used to inscribe gratings consisting of 20 to 40 lines, each 1-microm wide, 100-microm long, and 5-microm apart, 100 mum below the tissue surface. Refractive index changes in the micromachined regions were calculated immediately and after 1 month of storage by measuring the intensity distribution of diffracted light when the gratings were irradiated with a 632.8-nm He-Ne laser.
Periodic gratings were created in the stromal layer of the corneas and the cortex of the lenses by adjusting the laser pulse energy until visible plasma luminescence and bubbles were no longer generated. The gratings had low scattering loss and could only be visualized using phase microscopy. Refractive index changes measured 0.005 +/- 0.001 to 0.01 +/- 0.001 in corneal tissue and 0.015 +/- 0.001 to 0.021 +/- 0.001 in the lenses. The gratings and refractive index changes were preserved after storing the micromachined corneas and lenses for 1 month.
These pilot experiments demonstrate a novel application of low-pulse-energy, MHz femtosecond lasers in modifying the refractive index of transparent ocular tissues without apparent tissue destruction. Although it remains to be verified in living tissues, the stability of this effect suggests that the observed modifications are due to long-term molecular and/or structural changes.

0 Followers
 · 
109 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: A precision, large stroke (nearly 1 cm) scanning system was designed, built, and calibrated for micromachining of ophthalmic materials including hydrogels and cornea (excised and in vivo). This system comprises a flexure stage with an attached objective on stacked vertical and horizontal translation stages. This paper outlines the design process leading to our most current version including the specifications that were used in the design and the drawbacks of other methods that were previously used. Initial measurements of the current version are also given. The current flexure was measured to have a 27 Hz natural frequency with no load.
    Review of Scientific Instruments 06/2014; 85(6):065107-065107-11. DOI:10.1063/1.4880727 · 1.58 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Purpose: To determine the efficacy of intra-tissue refractive index shaping ("IRIS") using 400 nm femtosecond laser pulses (blue light) for writing refractive structures directly into live cat corneas in vivo, and to assess the longevity of these structures in the eyes of living cats. Methods: Four eyes from two adult cats underwent Blue-IRIS. Light at 400 nm with 100 femtosecond (fs) pulses were tightly focused into the corneal stroma of each eye at an 80 MHz repetition rate. These pulses locally increased the refractive index of the corneal stroma via an endogenous, two-photon absorption process and were used to inscribe 3-layered, gradient index patterns into the cat corneas. The optical effects of the patterns were then tracked using optical coherence tomography (OCT) and Shack-Hartmann wavefront sensing. Results: Blue-IRIS patterns locally changed ocular cylinder by -1.4 ± 0.3 D, defocus by 2.0 ± 0.5 D, and higher order root mean square (HORMS) by 0.31 ± 0.04 μm at 1 month post-IRIS, without significant changes in corneal thickness or curvature. Refractive changes were maintained for the duration they were tracked, 12 months post-IRIS in one eye, and just over 3 months in the remaining 3 eyes. Conclusions: Blue-IRIS can be used to inscribe refractive structures into live cat cornea in vivo that are stable for at least 12 months, and are not associated with significant alterations in corneal thicknesses or radii of curvature. This result is a critical step towards establishing Blue-IRIS as a promising technique for non-invasive vision correction.
    Investigative Ophthalmology &amp Visual Science 07/2014; 55(7). DOI:10.1167/iovs.14-14373 · 3.66 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Femtosecond (fs) lasers for LASIK flap creation have increased their market share from 0 percent in 2001 to over 55 percent in 2010. Now, in cataract surgery a similar revolution may take place. What makes this technology so special and which new areas of surgery will furthermore be entered in the future? This review elucidates the most important technical features of the fs lasers and interaction processes with the tissue. This understanding is intended to provide a better overview of the current market situation on the one hand. In addition, it can be estimated what kind of short- and long-term development this technology could take in the future.
    Klinische Monatsblätter für Augenheilkunde 12/2013; 230(12):1207-12. DOI:10.1055/s-0033-1351058 · 0.67 Impact Factor