Clinical Determinants of Radiation Dose in Percutaneous Coronary Interventional Procedures

Division of Cardiovascular Diseases, Mayo Clinic, Rochester, Minnesota 55905, USA.
JACC. Cardiovascular Interventions (Impact Factor: 7.44). 03/2011; 4(3):336-43. DOI: 10.1016/j.jcin.2010.10.014
Source: PubMed

ABSTRACT The objectives of this work were to establish the primary clinical determinants of patient radiation dose associated with percutaneous coronary interventional (PCI) and to identify opportunities for dose reduction.
Use of X-ray imaging and associated radiation dose is a necessary part of PCI. Potential adverse consequences of radiation dose include skin radiation injury and predicted increase in lifetime cancer risk.
Cumulative skin dose (CSD) (measured in gray [Gy] units) was selected as a measurement of patient radiation burden. Several patient-, disease-, and treatment-related variables, including 15 performing physicians, were analyzed in a multiple linear regression statistical model with cumulative skin dose CSD as the primary end point. The model results provide an estimate of the relative CSD increase (decrease) attributable to each variable.
Percutaneous coronary interventions performed on 1,287 male and 540 female patients were included. Median patient age was 68.6 years, median body mass index was 29.7 kg/m(2), and median weight was 88 kg. Median CSD was 1.64 Gy per procedure for male and 1.15 Gy for female patients. Increasing body mass index, patient sex, lesion complexity, lesion location, and performing physician were significantly associated with CSD. Physicians who performed more procedures were associated with lower CSD.
Several primary determinants of patient radiation dose during PCI were identified. Along with physician development of radiation-sparing methods and skills, pre-procedure dose planning is proposed to help minimize radiation dose for PCI.

0 Followers
 · 
83 Views
  • Circulation Cardiovascular Interventions 08/2014; 7(4):425-8. DOI:10.1161/CIRCINTERVENTIONS.114.001815 · 6.98 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Background-Reducing digital pulse rates (PR) are known to reduce total energy during invasive cardiovascular procedures, which likely has benefits for patients and staff. Physicians may be reluctant to reduce these parameters because they fear a decline in image quality that could affect procedural outcomes. We sought to assess the effect of default rates of fluoroscopy (Fluoro) and CINE-acquisition (CINE) on total x-ray dose and image quality during invasive cardiovascular procedures. Methods and Results-We retrospectively reviewed procedures done with 2 default PRs: a standard dose cohort (PR, 15 for Fluoro and CINE), and a reduced dose cohort (PR, 10 for Fluoro and CINE). Total x-ray dose, Fluoro time, and contrast use were compared between groups. A blinded angiographic image quality assessment was then performed using an objective 10-point angiographic quality score. There were no significant differences between cohorts for fluoroscopy time or contrast use. The reduced dose cohort has a significant reduction in mean total x-ray dose (PR 15, 1763.1 mGy; PR 10, 1179.1 mGy; P<0.0001). When adjusted for potential confounders, a 38% reduction in total x-ray dose was identified (P<0.0001). There was no difference in adjusted angiographic quality score between the cohorts (PR 15, 7.90; PR 10, 8.00; P=0.67), indicating no decline in image quality with PR reduction. Conclusions-Reducing default PRs during invasive cardiovascular procedures yields large and significant reductions in total x-ray energy with no decline in angiographic image quality.
    Circulation Cardiovascular Interventions 07/2014; 7(4). DOI:10.1161/CIRCINTERVENTIONS.114.001479 · 6.98 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Objectives The aim of this study was to define age-stratified, procedure-specific benchmark radiation dose levels during interventional catheterization for congenital heart disease. Background There is a paucity of published literature with regard to radiation dose levels during catheterization for congenital heart disease. Obtaining benchmark radiation data is essential for assessing the impact of quality improvement initiatives for radiation safety. Methods Data were obtained retrospectively from 7 laboratories participating in the Congenital Cardiac Catheterization Project on Outcomes collaborative. Total air kerma, dose area product, and total fluoroscopy time were obtained for the following procedures: 1) patent ductus arteriosus closure; 2) atrial septal defect closure; 3) pulmonary valvuloplasty; 4) aortic valvuloplasty; 5) treatment of coarctation of aorta; and 6) transcatheter pulmonary valve placement. Results Between January 2009 and July 2013, 2,713 cases were identified. Radiation dose benchmarks are presented including median, 75th percentile, and 95th percentile. Radiation doses varied widely between age groups and procedure types. Radiation exposure was lowest in patent ductus arteriosus closure and highest in transcatheter pulmonary valve placement. Total fluoroscopy time was a poor marker of radiation exposure and did not correlate well with total air kerma and dose area product. Conclusions This study presents age-stratified radiation dose values for 6 common congenital heart interventional catheterization procedures. Fluoroscopy time alone is not an adequate measure for monitoring radiation exposure. These values will be used as baseline for measuring the effectiveness of future quality improvement activities by the Congenital Cardiac Catheterization Project on Outcomes collaborative.
    JACC Cardiovascular Interventions 09/2014; 7(9):1060–1069. DOI:10.1016/j.jcin.2014.04.013 · 7.44 Impact Factor