Nine Orders of Magnitude Dynamic Range: Picomolar to Millimolar Concentration Measurement in Capillary Electrophoresis with Laser Induced Fluorescence Detection Employing Cascaded Avalanche Photodiode Photon Counters

Department of Chemistry and Biochemistry, University of Notre Dame, Notre Dame, Indiana 46556, USA.
Analytical Chemistry (Impact Factor: 5.64). 03/2011; 83(7):2748-53. DOI: 10.1021/ac103374x
Source: PubMed


The dynamic range of capillary electrophoresis analysis is ultimately limited by molecular shot noise at low concentrations and by concentration-induced band broadening at high concentrations. We report a system that approaches these fundamental limits. A laser-induced fluorescence detector is reported that employs a cascade of four fiber-optic beam splitters connected in series to generate a primary signal and four attenuated signals, each monitored by a single-photon counting avalanche photodiode. Appropriate scaling of the signals from the five photodiodes produces a linear optical calibration curve for 5-carboxyl-tetramethylrhodamine from the concentration detection limit of 1 pM to the upper limit of 1 mM. Mass detection limits are 120 yoctomoles (70 molecules) injected into the instrument. The very-wide dynamic range instrument was used to study the metabolic products of the fluorescently labeled glycosphingolipid tetramethylrhodamine labeled GM1 (GM1-TMR) produced by single cells isolated from the rat cerebellum.

123 Reads
    • "Laser-induced fluorescence (LIF) is the most sensitive detection technique used with CE (Pentoney and Sweedler, 1997); it can not only detect but quantify neurotransmitters in minute volumes of sample based on their retention time and area under the peak curve, and can be used to confirm analyte identity by spectral characteristics relative to known standards (Wise and Shear, 2006; Hatcher et al., 2008; Fossat et al., 2012). A dynamic range of nine orders of magnitude and 120 yoctomole sensitivity have been reported for glycosphingolipid metabolites (Dada et al., 2011). Applications of this approach to the analysis of single cells and smallvolume biological samples have been reviewed elsewhere (Szoko and Tabi, 2010; Lin et al., 2011), with several examples highlighted in the following sections of this article. "
    [Show abstract] [Hide abstract]
    ABSTRACT: Modern science is characterized by integration and synergy between research fields. Accordingly, as technological advances allow new and more ambitious quests in scientific inquiry, numerous analytical and engineering techniques have become useful tools in biological research. The focus of this review is on cutting edge technologies that aid direct measurement of bioactive compounds in the nervous system to facilitate fundamental research, diagnostics, and drug discovery. We discuss challenges associated with measurement of cell to cell signaling molecules in the nervous system, and advocate for a decrease of sample volumes to the nanoliter volume regime for improved analysis outcomes. We highlight effective approaches for the collection, separation, and detection of such small-volume samples, present strategies for targeted and discovery-oriented research, and describe the required technology advances that will empower future translational science.Neuropsychopharmacology accepted article preview online, 10 June 2013; doi:10.1038/npp.2013.145.
    Neuropsychopharmacology: official publication of the American College of Neuropsychopharmacology 06/2013; 39(1). DOI:10.1038/npp.2013.145 · 7.05 Impact Factor
  • Source
    • "Several attempts have been made to extend the usable range of photon counting out of the conventional Poisson counting limit. The most common approaches usually rely on diminishing the light that reaches the detector, either by attenuating the total signal or by dividing it across multiple detectors [6,7]. Unfortunately, all such methods suffer SNR losses in the high photon flux regime. "
    [Show abstract] [Hide abstract]
    ABSTRACT: Data from photomultiplier tubes are typically analyzed using either counting or averaging techniques, which are most accurate in the dim and bright signal limits, respectively. A statistical means of adjoining these two techniques is presented by recovering the Poisson parameter from averaged data and relating it to the statistics of binomial counting from Kissick et al. [Anal. Chem. 82, 10129 (2010)]. The point at which binomial photon counting and averaging have equal signal to noise ratios is derived. Adjoining these two techniques generates signal to noise ratios at 87% to approaching 100% of theoretical maximum across the full dynamic range of the photomultiplier tube used. The technique is demonstrated in a second harmonic generation microscope.
    Optics Express 04/2012; 20(9):10406-15. DOI:10.1364/OE.20.010406 · 3.49 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Organic molecules are a significant and highly varied component of atmospheric aerosols. Measurement of aerosol composition and improvements in our understanding of the complex chemistry involved in their formation and aging are being aided by innovations in soft ionization aerosol MS. (To listen to a podcast about this feature, please go to the Analytical Chemistry multimedia page at
    Analytical Chemistry 04/2011; 83(7):2409-15. DOI:10.1021/ac102737k · 5.64 Impact Factor