A multichannel time-domain scanning fluorescence mammograph: performance assessment and first in vivo results.

Physikalisch-Technische Bundesanstalt (PTB), Berlin, Germany.
The Review of scientific instruments (Impact Factor: 1.52). 02/2011; 82(2):024302. DOI: 10.1063/1.3543820
Source: PubMed

ABSTRACT We present a scanning time-domain fluorescence mammograph capable to image the distribution of a fluorescent contrast agent within a female breast, slightly compressed between two parallel glass plates, with high sensitivity. Fluorescence of the contrast agent is excited using a near infrared picosecond diode laser module. Four additional picosecond diode lasers with emission wavelengths between 660 and 1066 nm allow to measure the intrinsic optical properties of the breast tissue. By synchronously moving a source fiber and seven detection fiber bundles across the breast, distributions of times of flight of photons are recorded simultaneously for selected source-detector combinations in transmission and reflection geometry either at the fluorescence wavelength or at the selected laser wavelengths. To evaluate the performance of the mammograph, we used breastlike rectangular phantoms comprising fluorescent and absorbing objects using the fluorescent dye Omocyanine as contrast agent excited at 735 nm. We compare two-dimensional imaging of the phantom based on transmission and reflection data. Furthermore, we developed an improved tomosynthesis algorithm which permits three-dimensional reconstruction of fluorescence and absorption properties of lesions with good spatial resolution. For illustration, we present fluorescence mammograms of one patient recorded 30 min after administration of the contrast agent indocyanine green showing the carcinoma at high contrast originating from fluorescence of the extravasated dye, excited at 780 nm.

0 Bookmarks
 · 
89 Views
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: In the present contribution we investigate the images of CW diffusely reflected light for a point-like source, registered by a CCD camera imaging a turbid medium containing an absorbing lesion. We show that detection of μa variations (absorption anomalies) is achieved if images are normalized to background intensity. A theoretical analysis based on the diffusion approximation is presented to investigate the sensitivity and the limitations of our proposal and a novel procedure to find the location of the inclusions in 3D is given and tested. An analysis of the noise and its influence on the detection capabilities of our proposal is provided. Experimental results on phantoms are also given, supporting the proposed approach.
    Biomedical Optics Express 05/2014; 5(5):1336-54. · 3.18 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: To evaluate if clinical fluorescence imaging of IRDye800CW is feasible on our fluorescence optical mammography system by estimating detection limits assessed by breast-cancer-simulating phantom experiments. Phantoms (2.1 cm(3), 0.9 cm(3)) with IRDye800CW concentrations of 0.5 to 120 nM were suspended in a 550 cm(3) measurement cup containing 507 surface-mounted source and detector fibers. The cup was filled with optical matching fluid containing IRDye800CW concentrations of 0, 5, 10, or 20 nM. Tomographic fluorescence images were acquired by exciting IRDye800CW at 730 nm; wavelengths above 750 nm were filtered. Signal intensities were calculated over a volume of interest corresponding to the size and location of the phantom in the reconstructed images. Correlations (R(2)) were calculated, and detection limits with associated upper 95% prediction interval were estimated. Between-day reproducibility was assessed with intraclass correlation coefficients (ICC). Fluorescent intensities were strongly correlated with phantom IRDye800CW concentrations (R(2)0.983 to 0.999). IRDye800CW detection limits ranged from 0.14 to 2.46 nM (upper 95% prediction limit 4.63 to 18.63 nM). ICC ranged from 0.88 to 1.00. The estimated detection limits for IRDye800CW were in the low-nanomolar range. These results support the start of clinical trials to evaluate the fluorescence optical mammography system using IRDye800CW labeled breast cancer targeting ligands.
    Journal of Biomedical Optics 07/2012; 17(7):076022. · 2.88 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Recent advances in the use of diffuse optical techniques for monitoring the hemodynamic, metabolic and physiological signatures of the neoadjuvant breast cancer therapy effectiveness is critically reviewed. An extensive discussion of the state-of-theart diffuse optical mammography is presented alongside a discussion of the current approaches to breast cancer therapies. Overall, the diffuse optics field is growing rapidly with a great deal of promise to fill an important niche in the current approaches to monitor, predict and personalize neoadjuvant breast cancer therapies.
    IEEE Journal of Selected Topics in Quantum Electronics 01/2012; 18(4):1367-1386. · 4.08 Impact Factor