Article

Acoustic cardiac triggering: a practical solution for synchronization and gating of cardiovascular magnetic resonance at 7 Tesla.

Berlin Ultrahigh Field Facility, Max-Delbrueck Center for Molecular Medicine, Berlin, Germany.
Journal of Cardiovascular Magnetic Resonance (impact factor: 3.72). 11/2010; 12:67. DOI:10.1186/1532-429X-12-67 pp.67
Source: PubMed

ABSTRACT To demonstrate the applicability of acoustic cardiac triggering (ACT) for imaging of the heart at ultrahigh magnetic fields (7.0 T) by comparing phonocardiogram, conventional vector electrocardiogram (ECG) and traditional pulse oximetry (POX) triggered 2D CINE acquisitions together with (i) a qualitative image quality analysis, (ii) an assessment of the left ventricular function parameter and (iii) an examination of trigger reliability and trigger detection variance derived from the signal waveforms.
ECG was susceptible to severe distortions at 7.0 T. POX and ACT provided waveforms free of interferences from electromagnetic fields or from magneto-hydrodynamic effects. Frequent R-wave mis-registration occurred in ECG-triggered acquisitions with a failure rate of up to 30% resulting in cardiac motion induced artifacts. ACT and POX triggering produced images free of cardiac motion artefacts. ECG showed a severe jitter in the R-wave detection. POX also showed a trigger jitter of approximately Δt = 72 ms which is equivalent to two cardiac phases. ACT showed a jitter of approximately Δt = 5 ms only. ECG waveforms revealed a standard deviation for the cardiac trigger offset larger than that observed for ACT or POX waveforms.Image quality assessment showed that ACT substantially improved image quality as compared to ECG (image quality score at end-diastole: ECG = 1.7 ± 0.5, ACT = 2.4 ± 0.5, p = 0.04) while the comparison between ECG vs. POX gated acquisitions showed no significant differences in image quality (image quality score: ECG = 1.7 ± 0.5, POX = 2.0 ± 0.5, p = 0.34).
The applicability of acoustic triggering for cardiac CINE imaging at 7.0 T was demonstrated. ACT's trigger reliability and fidelity are superior to that of ECG and POX. ACT promises to be beneficial for cardiovascular magnetic resonance at ultra-high field strengths including 7.0 T.

0 0
 · 
0 Bookmarks
 · 
39 Views
  • Article: Cardiac imaging using gated magnetic resonance.
    [show abstract] [hide abstract]
    ABSTRACT: To overcome the limitations of magnetic resonance (MR) cardiac imaging using nongated data acquisition, three methods for acquiring a gating signal, which could be applied in the presence of a magnetic field, were tested: an air-filled plethysmograph, a laser-Doppler capillary perfusion flowmeter, and an electrocardiographic gating device. The gating signal was used for timing of MR imaging sequences (IS). Application of each gating method yielded significant improvements in structural MR image resolution of the beating heart, although with both plethysmography and laser-Doppler velocimetry it was difficult to obtain cardiac images from the early portion of the cardiac cycle due to an intrinsic delay between the ECG R wave and peripheral detection of the gating signal. Variations in the temporal relationship between the R wave and plethysmographic and laser-Doppler signals produced inconsistencies in the timing of IS. Since the ECG signal is virtually free of these problems, the preferable gating technique is IS synchronization with an electrocardiogram. The gated images acquired with this method provide sharp definition of internal cardiac morphology and can be temporarily referenced to end diastole and end systole or intermediate points.
    Radiology 02/1984; 150(1):121-7. · 5.73 Impact Factor
  • Source
    Article: Fiber-optic stethoscope: a cardiac monitoring and gating system for magnetic resonance microscopy.
    [show abstract] [hide abstract]
    ABSTRACT: A fundamental problem associated with using the conventional electrocardiograph (ECG) to monitor a subject's cardiac activity during magnetic resonance imaging (MRI) is the distortion of the ECG due to electromagnetic interference. This problem is particularly pronounced in MR microscopy (MRI of small animals at microscopic resolutions (< 0.03 mm(3))) because the strong, rapidly-switching magnetic field gradients induce artifacts in the animal's ECG that often mimic electrophysiologic activity, impairing the use of the ECG for cardiac monitoring and gating purposes. The fiber-optic stethoscope system offers a novel approach to measuring cardiac activity that, unlike the ECG, is immune to electromagnetic effects. The fiber-optic stethoscope is perorally inserted into the esophagus of small animals to optically detect pulsatile compression of the esophageal wall. The optical system is shown to provide a robust cardiac monitoring and gating signal in rats and mice during routine cardiac MR microscopy.
    Magnetic Resonance in Medicine 03/2002; 47(2):314-21. · 2.96 Impact Factor
  • Article: Cardiovascular MR imaging: pressure-gating using the arterial pressure signal from a conventional ferromagnetic micromanometer-tip catheter.
    [show abstract] [hide abstract]
    ABSTRACT: We used the arterial pressure wave as a reliable trigger to synchronize MR imaging to the cardiac cycle. Pressure readings and MR images remained undistorted using a conventional ferromagnetic micromanometer-tip catheter. As a safe alternative to triggering to ventricular pressure, this approach allows varying the time relation between the trigger and the cardiac cycle while maintaining the advantages of pressure-gating.
    Magnetic Resonance Imaging 02/1994; 12(3):531-4. · 1.99 Impact Factor

Full-text (2 Sources)

View
0 Downloads
Available from
29 Jan 2013

Keywords

acoustic cardiac
 
ACT promises
 
cardiac CINE imaging
 
cardiac motion artefacts
 
cardiovascular magnetic resonance
 
conventional vector electrocardiogram
 
ECG waveforms
 
electromagnetic fields
 
Frequent R-wave mis-registration
 
image quality score
 
left ventricular function parameter
 
magneto-hydrodynamic effects
 
POX waveforms.Image quality assessment
 
qualitative image quality analysis
 
R-wave detection
 
severe distortions
 
signal waveforms
 
standard deviation
 
traditional pulse oximetry
 
ultrahigh magnetic fields