Pyrolytic graphite foam: a passive magnetic susceptibility matching material.

Berkeley/UCSF Bioengineering Joint Graduate Group, University of California, Berkeley, California 94720, USA.
Journal of Magnetic Resonance Imaging (Impact Factor: 2.79). 09/2010; 32(3):684-91. DOI: 10.1002/jmri.22270
Source: PubMed

ABSTRACT To evaluate a novel soft, lightweight cushion that can match the magnetic susceptibility of human tissue. The magnetic susceptibility difference between air and tissue produces field inhomogeneities in the B(0) field, which leads to susceptibility artifacts in magnetic resonance imaging (MRI) studies.
Pyrolytic graphite (PG) microparticles were uniformly embedded into a foam cushion to reduce or eliminate field inhomogeneities at accessible air and tissue interfaces. 3T MR images and field maps of an air/water/PG foam phantom were acquired. Q measurements on a 4T tuned head coil and pulse sequence heating tests at 3T were also performed.
The PG foam improved susceptibility matching, reduced the field perturbations in phantoms, does not heat, and is nonconductive.
The susceptibility matched PG foam is lightweight, safe for patient use, adds no noise or MRI artifacts, is compatible with radiofrequency coil arrays, and improves B(0) homogeneity, which enables more robust MR studies.

0 Followers
 · 
157 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: To demonstrate dynamic slice-dependent shim update as a simple method to reduce susceptibility-induced B0 inhomogeneity and associated pixel shift artifacts in diffusion-weighted echo planar imaging (DW-EPI) in 3 T breast imaging. Dynamic slice-dependent update of linear shim and center frequency was implemented in a dual-echo B0 mapping sequence and a DW-EPI sequence. Multi-slice axial B0 maps and diffusion-weighted images were obtained from four volunteers with both conventional and dynamic shim methods. The two shim methods were compared in terms of B0 homogeneity and EPI pixel shift artifacts. In all volunteers the B0 maps showed significantly improved homogeneity; the left-right asymmetry was reduced by 79% and within-slice B0 standard deviation was reduced by 20% on the average. The improvements were better than what was previously reported for conventional (static) third-order shim in bilateral breast. Anatomy-referenced apparent diffusion coefficient (ADC) maps showed reduced overall image registration error obtainable with dynamic shim. Dynamic shim is an effective method to improve B0 shimming and DW-EPI image quality in 3 T bilateral breast imaging. Magn Reson Med, 2013. © 2013 Wiley Periodicals, Inc.
    Magnetic Resonance in Medicine 05/2014; 71(5). DOI:10.1002/mrm.24824 · 3.40 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: We propose a design of a integrated system of optic fibers and a superconducting surface coil. A MRI compatible support fixed the optic fiber and the coil, and the optic fibers installed through the air gap. In order to compensate the periodic errors of the hemoglobin concentration estimation produced by the respiration and heart rate of subjects, we propose the generalized least square method to acquire the adaptive estimation, which considers the noise variance of the temporal optical measurements as a time-varying model. The vivo experiments that the optical-sensing simultaneously working with MRI scans are conducted to characterize the performance of this approach.
    2014 IEEE Workshop on Electronics, Computer and Applications (IWECA); 05/2014
  • [Show abstract] [Hide abstract]
    ABSTRACT: To demonstrate improved frequency selective fat suppression in MRI using a magnetic susceptibility matching foam by reducing B0 inhomogeneities induced within the body by air-tissue interfaces. Flexible pyrolytic graphite (PG) composite foam was tailored to match the magnetic susceptibility of human tissue and was shaped to surround the cervical spine region. Field maps and frequency selective fat suppressed T1 -weighted FLASH images were acquired at 3 Tesla in both phantoms and six healthy necks. B0 field uniformity was shimmed to a target critical threshold of 1 ppm for fat suppression. The percentage of voxels in the phantom that did not achieve the critical threshold was reduced from 64% without the PG foam to only 1% with the foam. A similar decrease from 16% to 2% was observed in the in vivo region of interest. PG foam improved B0 field uniformity by moving air-tissue field gradients outside of the neck where they cannot cause MRI artifacts. The PG foams consistently mitigated signal dropout, improved overall SNR, and enabled more robust frequency selective fat suppression.J. Magn. Reson. Imaging 2014. © 2014 Wiley Periodicals, Inc.
    Journal of Magnetic Resonance Imaging 03/2015; 41(3). DOI:10.1002/jmri.24581 · 2.79 Impact Factor