W-band frequency-swept EPR.

National Biomedical EPR Center, Department of Biophysics, Medical College of Wisconsin, Milwaukee, WI 53226, USA.
Journal of Magnetic Resonance (Impact Factor: 2.3). 07/2010; 205(1):93-101. DOI: 10.1016/j.jmr.2010.04.005
Source: PubMed

ABSTRACT This paper describes a novel experiment on nitroxide radical spin labels using a multiarm EPR W-band bridge with a loop-gap resonator (LGR). We demonstrate EPR spectroscopy of spin labels by linear sweep of the microwave frequency across the spectrum. The high bandwidth of the LGR, about 1 GHz between 3 dB points of the microwave resonance, makes this new experiment possible. A frequency-tunable yttrium iron garnet (YIG) oscillator provides sweep rates as high as 1.8x10(5) GHz/s, which corresponds to 6.3 kT/s in magnetic field-sweep units over a 44 MHz range. Two experimental domains were identified. In the first, linear frequency sweep rates were relatively slow, and pure absorption and pure dispersion spectra were obtained. This appears to be a practical mode of operation at the present level of technological development. The main advantage is the elimination of sinusoidal magnetic field modulation. In the second mode, the frequency is swept rapidly across a portion of the spectrum, and then the frequency sweep is stopped for a readout period; FID signals from a swept line oscillate at a frequency that is the difference between the spectral position of the line in frequency units and the readout position. If there is more than one line, oscillations are superimposed. The sweep rates using the YIG oscillator were too slow, and the portion of the spectrum too narrow to achieve the full EPR equivalent of Fourier transform (FT) NMR. The paper discusses technical advances required to reach this goal. The hypothesis that trapezoidal frequency sweep is an enabling technology for FT EPR is supported by this study.

0 Bookmarks
 · 
130 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: Recent advances in electron paramagnetic resonance (EPR) include capabilities for applications to areas as diverse as archeology, beer shelf life, biological structure, dosimetry, in vivo imaging, molecular magnets, and quantum computing. Enabling technologies include multifrequency continuous wave, pulsed, and rapid scan EPR. Interpretation is enhanced by increasingly powerful computational models.
    Journal of Magnetic Resonance 08/2012; 223:151-63. DOI:10.1016/j.jmr.2012.07.025 · 2.30 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: A resonated coil driver designed for rapid scan electron paramagnetic resonance (EPR) is described. Rapid scan EPR requires larger magnetic field coils and a more powerful magnetic field modulation than those designed for conventional continuous wave EPR. There are two advantages of a resonated coil driver: i) by resonating the scan coils the reactive components are removed, which allows the coils to be driven with much lower voltages, and ii) by resonating the coils a near-perfect sinusoidal scan is ensured. Performance of the coil driver is improved when Litz wire coils are used. With coil diameters between 3 and 3.5 inches, sweeps can be generated with widths up to about 80 G peak-to-peak and frequencies up to about 150 kHz, which permits sweep rates of more than 8 MG/s. The magnetic field is uniform over the sample length of a Bruker dielectric resonator. Enough detail has been presented for other laboratories toimplement this design to perform rapid scan EPR. © 2012 Wiley Periodicals, Inc. Concepts Magn Reson Part B (Magn Reson Engineering) 41B: 95–110, 2012
    Concepts in Magnetic Resonance Part B Magnetic Resonance Engineering 10/2012; 41B(4). DOI:10.1002/cmr.b.21222 · 0.71 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Arbitrary waveform generator based X-band EPR spectrometer.•Signal is down-converted to intermediate frequency near baseband.•EPR signal is digitized and converted to quadrature from single physical channel.•The method is demonstrated for rapid scan of 4 organic radicals.•The method is applicable for any periodic signal.
    Journal of Magnetic Resonance 10/2014; DOI:10.1016/j.jmr.2014.10.011 · 2.32 Impact Factor