Article

Simultaneous dual-radionuclide myocardial perfusion imaging with a solid-state dedicated cardiac camera.

Institute of Nuclear Medicine, University College London Hospitals NHS Trust, London, UK.
European Journal of Nuclear Medicine (impact factor: 4.53). 04/2010; 37(9):1710-21. DOI:10.1007/s00259-010-1441-1 pp.1710-21
Source: PubMed

ABSTRACT We compared simultaneous dual-radionuclide (DR) stress and rest myocardial perfusion imaging (MPI) with a novel solid-state cardiac camera and a conventional SPECT camera with separate stress and rest acquisitions.
Of 27 consecutive patients recruited, 24 (64.5+/-11.8 years of age, 16 men) were injected with 74 MBq of (201)Tl (rest) and 250 MBq (99m)Tc-MIBI (stress). Conventional MPI acquisition times for stress and rest are 21 min and 16 min, respectively. Rest (201)Tl for 6 min and simultaneous DR 15-min list mode gated scans were performed on a D-SPECT cardiac scanner. In 11 patients DR D-SPECT was performed first and in 13 patients conventional stress (99m)Tc-MIBI SPECT imaging was performed followed by DR D-SPECT. The DR D-SPECT data were processed using a spill-over and scatter correction method. DR D-SPECT images were compared with rest (201)Tl D-SPECT and with conventional SPECT images by visual analysis employing the 17-segment model and a five-point scale (0 normal, 4 absent) to calculate the summed stress and rest scores. Image quality was assessed on a four-point scale (1 poor, 4 very good) and gut activity was assessed on a four-point scale (0 none, 3 high).
Conventional MPI studies were abnormal at stress in 17 patients and at rest in 9 patients. In the 17 abnormal stress studies DR D-SPECT MPI showed 113 abnormal segments and conventional MPI showed 93 abnormal segments. In the nine abnormal rest studies DR D-SPECT showed 45 abnormal segments and conventional MPI showed 48 abnormal segments. The summed stress and rest scores on conventional SPECT and DR D-SPECT were highly correlated (r=0.9790 and 0.9694, respectively). The summed scores of rest (201)Tl D-SPECT and DR-DSPECT were also highly correlated (r=0.9968, p<0.0001 for all). In six patients stress perfusion defects were significantly larger on stress DR D-SPECT images, and five of these patients were imaged earlier by D-SPECT than by conventional SPECT.
Fast and high-quality simultaneous DR MPI is feasible with D-SPECT in a single imaging session with comparable diagnostic performance and image quality to conventional SPECT and to a separate rest (201)Tl D-SPECT acquisition.

0 0
 · 
0 Bookmarks
 · 
38 Views

Keywords

11 patients DR D-SPECT
 
17 patients
 
17-segment model
 
201)Tl D-SPECT acquisition
 
27 consecutive patients recruited
 
93 abnormal segments
 
Conventional MPI studies
 
conventional SPECT camera
 
conventional SPECT images
 
D-SPECT cardiac scanner
 
DR D-SPECT
 
DR D-SPECT images
 
five-point scale
 
four-point scale
 
nine abnormal rest studies DR D-SPECT
 
novel solid-state cardiac camera
 
simultaneous dual-radionuclide
 
single imaging session
 
stress DR D-SPECT images
 
visual analysis