Totally automatic definition of renal regions of interest from 99mTc-MAG3 renograms: validation in patients with normal kidneys and in patients with suspected renal obstruction.

Department of Radiology, Emory University School of Medicine, Atlanta, Georgia, USA.
Nuclear Medicine Communications (impact factor: 1.4). 02/2010; 31(5):366-74. DOI:10.1097/MNM.0b013e3283362aa3 pp.366-74
Source: PubMed

ABSTRACT An image-processing algorithm (AUTOROI) has been developed to totally automatically (or with manual assistance) detect whole-kidney contours and generate renal regions of interest (ROI) for the extraction of the quantitative measurements used in the interpretation of Tc-mercaptoacetyltriglycine (Tc-MAG3) renograms.
The 18-20th min dynamic frames post-MAG3 injection were used to automatically define boxes surrounding each kidney, which were then transposed to an early composite image for interpolative and directional background subtraction. Sobel operator and unsharp masking were applied for edge enhancement, and the resulting image histograms were equalized to better define poorly functioning kidneys. AUTOROI searched radially from the center of mass to define each kidney's ROI coordinates. AUTOROI was validated using MAG3 studies from 79 patients referred for suspected obstruction (79 left, 77 right kidneys) and 19 kidney donors with normal kidney function and no obstruction. Renal ROIs were manually defined by a nuclear medicine technologist with 20+ years of experience (reference standard) and an American Board of Nuclear Medicine certified physician. AUTOROI and physician ROIs were automatically compared with the reference standard to determine the border definition error.
AUTOROI totally automatically detected the renal borders in 89% (172 of 194) of the kidneys from the entire group of 98 patients. The 22 kidneys missed automatically were subsequently detected with the assistance of a single manually placed fiducial point demarcating the liver/kidney boundary. These 22 kidneys were shown to be associated with markedly reduced MAG3 clearance. The mean error of AUTOROI for all 194 kidneys was 6.66+/-3.77 and 7.31+/-4.52 mm for the left and right kidney, respectively. The physician's error was 6.78+/-2.42 and 6.65+/-2.05 mm for the left and right kidney, respectively. This error difference between AUTOROI and the physician was not statistically significant.
AUTOROI provides an objective and promising approach to automated renal ROI detection.

0 0
 · 
0 Bookmarks
 · 
16 Views

Keywords

border definition error
 
composite image
 
directional background subtraction
 
error difference
 
image-processing algorithm
 
kidney's ROI
 
liver/kidney boundary
 
MAG3 clearance
 
manual assistance
 
mean error
 
Nuclear Medicine certified physician
 
nuclear medicine technologist
 
physician's error
 
reference standard
 
renal regions
 
renal ROI detection
 
Renal ROIs
 
resulting image histograms
 
single manually
 
unsharp masking
 

Ernest V Garcia