Comparison of three instruments for measuring red blood cell aggregation.

Department of Physiology, Faculty of Medicine, Akdeniz University, Antalya, Turkey.
Clinical hemorheology and microcirculation (Impact Factor: 3.4). 01/2009; 43(4):283-98. DOI: 10.3233/CH-2009-1240
Source: PubMed

ABSTRACT The International Society for Clinical Hemorheology organized a workshop to compare three instruments for measuring RBC aggregation: LORCA, Myrenne Aggregometer and RheoScan-A. The Myrenne Aggregometer provides indices at stasis (M) and at low shear (M1), with four indices obtained with the LORCA and RheoScan-A: amplitude (AMP), half-time (T1/2), surface area (SA) above (LORCA) or below (RheoScan-A) the syllectogram, and the ratio (AI) of the area above (LORCA) or below (RheoScan-A) the syllectogram to total area (AI). Intra-assay reproducibility and biological variability were determined; also studied were RBC in diluted plasma and in 1% 500 kDa dextran, and 0.003% glutaradehyde (GA)-treated cells in plasma. All measurements were performed at 37 degrees C. Standardized difference values were used as a measure of power to detect differences. Salient results were: (1) intra-assay variations below 5% except for RheoScan-A AMP and SA; (2) biological variability greatest for T1/2 with other indices similar for the three devices; (3) all instruments detected progressive changes with plasma dilution; (4) the Myrenne and LORCA, but not the RheoScan-A, detected differences for cells in dextran; (5) GA-treatment significantly affected the LORCA (AMP, T1/2, SA, AI), the RheoScan-A (AMP, SA, AI) and the Myrenne M parameter. It is concluded that: (a) the LORCA, Myrenne and the RheoScan-A have acceptable precision and suitable power for detecting reduced aggregation due to plasma dilution; (b) greatly enhanced RBC aggregation may not be sensed by the RheoScan-A while the Myrenne M1 index may be insensitive to minor increases of cell rigidity; (c) future studies should define each instrument's useful range for detecting RBC aggregation.

2 Bookmarks
 · 
193 Views
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: We sought to investigate the influence of the extent of myocardial injury on left ventricular (LV) systolic and diastolic function in patients after reperfused acute myocardial infarction (AMI). Thirty-eight reperfused AMI patients underwent cardiac magnetic resonance (CMR) imaging after percutaneous coronary revascularization. The extent of myocardial edema and scarring were assessed by T2 weighted imaging and late gadolinium enhancement (LGE) imaging, respectively. Within a day of CMR, echocardiography was done. Using 2D speckle tracking analysis, LV longitudinal, circumferential strain, and twist were measured. Extent of LGE were significantly correlated with LV systolic functional indices such as ejection fraction (r = -0.57, p < 0.001), regional wall motion score index (r = 0.52, p = 0.001), and global longitudinal strain (r = 0.56, p < 0.001). The diastolic functional indices significantly correlated with age (r = -0.64, p < 0.001), LV twist (r = -0.39, p = 0.02), average non-infarcted myocardial circumferential strain (r = -0.52, p = 0.001), and LV end-diastolic wall stress index (r = -0.47, p = 0.003 with e') but not or weakly with extent of LGE. In multivariate analysis, age and non-infarcted myocardial circumferential strain independently correlated with diastolic functional indices rather than extent of injury. In patients with timely reperfused AMI, not only extent of myocardial injury but also age and non-infarcted myocardial function were more significantly related to LV chamber diastolic function.
    Cardiovascular Ultrasound 02/2014; 12(1):6. · 1.32 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: A new method is described in this paper that allows measurement of red blood cell (RBC) aggregation indexes in disposable glass tubes within minutes. Light transmission through the RBC suspension filled into a microhematocrit capillary at stasis is recorded during RBC aggregation; a novel method assures an initial dispersion of aggregates in the capillary. The resulting light transmittance-time data are analyzed to calculate various parameters. Measurement of erythrocyte sedimentation rate (ESR) and RBC aggregation using well established methods and the newly developed capillary tube aggregometer in blood samples with a wide range of RBC aggregation indicated significant correlations between these parameters. Additionally, light transmittance during complete disaggregation allows estimating hematocrit, thereby enabling hematocrit correction of the measured and calculated parameters. The newly developed capillary tube RBC aggregometer is suitable for use as a method to rapidly monitor disease activity and the acute phase response, especially at the point-of-care (e.g., health care facilities, physician's office) and for field studies.
    Clinical hemorheology and microcirculation 01/2011; 47(4):295-305. · 3.40 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: The reversible aggregation of red blood cells (RBC) is of current basic science and clinical interest. Using a flow channel and light transmittance (LT) through RBC suspensions, we have examined the effects of wavelength (500 to 900 nm) on the static and dynamic aspects of RBC aggregation for normal blood and suspensions with reduced or enhanced aggregation; the effects of oxygenation were also explored. Salient observations include: 1. significant effects of wavelength on aggregation parameters reflecting the extent of aggregation (i.e., number of RBC per aggregate); 2. no significant effects of wavelength on parameters reflecting the time course of RBC aggregation; 3. a prominent influence of hemoglobin oxygen saturation on both extent and time-course related aggregation parameters measured at wavelengths less than 700 nm, but only on the time-course at 800 nm; and 4. the power of parameters in detecting a given alteration of RBC aggregation is affected by wavelength, in general being greater at higher wavelengths. It is recommended that light sources with wavelengths around 800 nm be used in instruments for measuring RBC aggregation via LT.
    Journal of Biomedical Optics 11/2011; 16(11):117006. · 2.75 Impact Factor

Full-text (2 Sources)

Download
121 Downloads
Available from
Jun 4, 2014