Accuracy of postoperative end-tidal Pco2 measurements with mainstream and sidestream capnography in non-obese patients and in obese patients with and without obstructive sleep apnea.

Department of Anesthesiology and Perioperative Medicine, University of Louisville, USA.
Anesthesiology (Impact Factor: 6.17). 09/2009; 111(3):609-15. DOI: 10.1097/ALN.0b013e3181b060b6
Source: PubMed

ABSTRACT Obtaining accurate end-tidal carbon dioxide pressure measurements via nasal cannula poses difficulties in postanesthesia patients who are mouth breathers, including those who are obese and those with obstructive sleep apnea (OSA); a nasal cannula with an oral guide may improve measurement accuracy in these patients. The authors evaluated the accuracy of a mainstream capnometer with an oral guide nasal cannula and a sidestream capnometer with a nasal cannula that did or did not incorporate an oral guide in spontaneously breathing non-obese patients and obese patients with and without OSA during recovery from general anesthesia.
The study enrolled 20 non-obese patients (body mass index less than 30 kg/m) without OSA, 20 obese patients (body mass index greater than 35 kg/m) without OSA, and 20 obese patients with OSA. End-tidal carbon dioxide pressure was measured by using three capnometer/cannula combinations (oxygen at 4 l/min): (1) a mainstream capnometer with oral guide nasal cannula, (2) a sidestream capnometer with a nasal cannula that included an oral guide, and (3) a sidestream capnometer with a standard nasal cannula. Arterial carbon dioxide partial pressure was determined simultaneously. The major outcome was the arterial-to-end-tidal partial pressure difference with each combination.
In non-obese patients, arterial-to-end-tidal pressure difference was 3.0 +/- 2.6 (mean +/- SD) mmHg with the mainstream capnometer, 4.9 +/- 2.3 mmHg with the sidestream capnometer and oral guide cannula, and 7.1 +/- 3.5 mmHg with the sidestream capnometer and a standard cannula (P < 0.05). In obese non-OSA patients, it was 3.9 +/- 2.6 mmHg, 6.4 +/- 3.1 mmHg, and 8.1 +/- 5.0 mmHg, respectively (P < 0.05). In obese OSA patients, it was 4.0 +/- 3.1 mmHg, 6.3 +/- 3.2 mmHg, and 8.3 +/- 4.6 mmHg, respectively (P < 0.05).
Mainstream capnometry performed best, and an oral guide improved the performance of sidestream capnometry. Accuracy in non-obese and obese patients, with and without OSA, was similar.

1 Bookmark
 · 
124 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: Current clinical guidelines recommend capnography as one of the best non-invasive methods to assess adequacy of ventilation in the non-intubated patient. Alveolar hypoventilation or respiratory depression is a serious event that occurs in a variety of clinical settings where patients receive sedatives and opioids. With the large number of procedures performed outside the operating room under the effects of sedatives and the increased use of patient-controlled analgesia, the need for capnography for monitoring has dramatically increased. Despite the succesful use of capnography to monitor ventilation in the operating room over several decades, other clinical areas have been very slow adapters of the technology and still rely heavily upon pulse oximetry to detect hypoventilation. This article reviews the most current evidence for using capnography in the non-intubated patient and summarizes the results of outcome measures reported in recent clinical trials. Capnography should be routinely used for non-intubated patients at risk for respiratory depression, in particular those receiving supplemental oxygen.
    Expert Review of Respiratory Medicine 07/2014;
  • [Show abstract] [Hide abstract]
    ABSTRACT: Respiratory depression, a potentially serious complication after general anaesthesia, can be detected promptly by close monitoring of both oxygen saturation and respiratory rate. Obese patients have morphological changes that may impair the reliability of monitoring devices. In this study, respiration rate was simultaneously recorded every second for up to 60 min using a computer in 30 adult obese patients (body mass index ≥ 35 kg.m−2), by three methods: acoustic; thoracic impedance; and capnometry via a facemask (Capnomask®, reference method). Of the 99 771 data triplets collected, only 85 520 (86%) were included; 12 021 (84%) were not studied due to failure of capnometry and 2240 (16%) due to failure of the acoustic method. Compared with capnometry, bias was similar using both the acoustic method and impedance (−0.3 bpm vs −0.6 bpm, respectively, p = 0.09), but limits of agreement were narrower for the acoustic method (±3.5 bpm vs ±5.3 bpm, respectively, p = 0.0008). The proportion of respiration rate values obtained with the acoustic method and impedance that differed by at least 10% or 20% for more than 15 s were 11% vs 23% and 2% vs 6%, respectively (p = 0.0009 for both comparisons). The acoustic sensor was well tolerated, while the facemask was pulled off on several occasions by four (13%) agitated patients. In obese patients requiring close monitoring of respiration rate, the acoustic method may be more precise than thoracic impedance and better tolerated than capnometry with a facemask.
    Anaesthesia 07/2014; · 3.85 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Capnometry is a method to measure carbon dioxide (CO2) in exhaled gas and its use during polysomnography (PSG) for diagnostic of sleep apnea-hypopnea syndrome is expanding. However, some problems exist for using capnometer in combination with other respiratory monitoring devices because capnometry requires additional sampling cannula or airway adapter attached to patients. To resolve these problems, we developed a novel mainstream capnometer system for PSG, which is designed to integrate multiple devices for measuring respiratory parameters. This system may provide comfortable and stable PSG including capnometry. We evaluated the basic performance of this system using a spontaneous breathing model. The result indicates that this newly developed system works adequately in PSG and moreover has superior characteristics of capnography signal and measurement stability against displacement of sensors, compared to conventional devices.
    08/2014; 2014:2139-43.

Full-text

Download
10 Downloads
Available from
Jun 5, 2014