Accuracy of Postoperative End-tidal P CO 2 Measurements with Mainstream and Sidestream Capnography in Non- obese Patients and in Obese Patients with and without Obstructive Sleep Apnea

Department of Anesthesiology and Perioperative Medicine, University of Louisville, USA.
Anesthesiology (Impact Factor: 5.88). 09/2009; 111(3):609-15. DOI: 10.1097/ALN.0b013e3181b060b6
Source: PubMed

ABSTRACT Obtaining accurate end-tidal carbon dioxide pressure measurements via nasal cannula poses difficulties in postanesthesia patients who are mouth breathers, including those who are obese and those with obstructive sleep apnea (OSA); a nasal cannula with an oral guide may improve measurement accuracy in these patients. The authors evaluated the accuracy of a mainstream capnometer with an oral guide nasal cannula and a sidestream capnometer with a nasal cannula that did or did not incorporate an oral guide in spontaneously breathing non-obese patients and obese patients with and without OSA during recovery from general anesthesia.
The study enrolled 20 non-obese patients (body mass index less than 30 kg/m) without OSA, 20 obese patients (body mass index greater than 35 kg/m) without OSA, and 20 obese patients with OSA. End-tidal carbon dioxide pressure was measured by using three capnometer/cannula combinations (oxygen at 4 l/min): (1) a mainstream capnometer with oral guide nasal cannula, (2) a sidestream capnometer with a nasal cannula that included an oral guide, and (3) a sidestream capnometer with a standard nasal cannula. Arterial carbon dioxide partial pressure was determined simultaneously. The major outcome was the arterial-to-end-tidal partial pressure difference with each combination.
In non-obese patients, arterial-to-end-tidal pressure difference was 3.0 +/- 2.6 (mean +/- SD) mmHg with the mainstream capnometer, 4.9 +/- 2.3 mmHg with the sidestream capnometer and oral guide cannula, and 7.1 +/- 3.5 mmHg with the sidestream capnometer and a standard cannula (P < 0.05). In obese non-OSA patients, it was 3.9 +/- 2.6 mmHg, 6.4 +/- 3.1 mmHg, and 8.1 +/- 5.0 mmHg, respectively (P < 0.05). In obese OSA patients, it was 4.0 +/- 3.1 mmHg, 6.3 +/- 3.2 mmHg, and 8.3 +/- 4.6 mmHg, respectively (P < 0.05).
Mainstream capnometry performed best, and an oral guide improved the performance of sidestream capnometry. Accuracy in non-obese and obese patients, with and without OSA, was similar.

Download full-text


Available from: Ozan Akca, May 05, 2014
1 Follower
38 Reads
  • [Show abstract] [Hide abstract]
    ABSTRACT: In this paper some criteria for testing model structures are used for choosing the best among a set of rival dynamical models of two metabolic processes. The considered criteria are based on the residual sum of squares, the final prediction error, the Akaike information criterion and the inverse of the information matrix. Linear time invariant structured models and input-output descriptions are proposed for the two considered metabolic processes, riphamicin and bilirubin kinetics, and maximum likelihood parameter estimates are obtained from input-output test data. A combined use of the Akaike criterion (or final prediction error) and of the inverse of the information matrix proved to be a very reliable tool in the metabolic model structure determination.
    Decision and Control including the Symposium on Adaptive Processes, 1979 18th IEEE Conference on; 01/1980
  • [Show abstract] [Hide abstract]
    ABSTRACT: This paper deals with a design of a digital servo controller for a robot manipulator with mechanical nonlinearities. The experimental system is a D.C. motor driven manipulator arm. A model to predict the nonlinear behavior exhibited by the experimental system and a nonlinear controller with Coulomb friction compensator are developed and tested. The effectiveness of the nonlinear controller is certified by designing a linear tracking controller based on this scheme. Experimental and simulated responses are given.
    Robotics and Automation. Proceedings. 1986 IEEE International Conference on; 05/1986
  • [Show abstract] [Hide abstract]
    ABSTRACT: Several devices have been proposed to monitor end-tidal carbon dioxide tension (Petco(2)) in spontaneously breathing patients; however, many have been reported to be inaccurate. We designed this study to investigate the accuracy of a balloon-tipped nasal catheter in measuring Petco(2) in nontracheally intubated, spontaneously breathing patients. The catheter was assembled using a 14-F rubber Foley catheter, a tracheal tube pilot balloon, and the plastic sheath from an 18-gauge needle. The catheter was connected to the sampling tube of a gas analyzer. Petco(2) and Paco(2) were determined simultaneously in 20 otherwise healthy postsurgical patients while receiving oxygen. The mean Petco(2) - Paco(2) difference was -4.4 +/- 1.6 (SD) mm Hg with a correlation coefficient r = +0.87 (P < 0.001). Our results suggest that a balloon-tipped nasal catheter can provide a simple, easy, and reliable method for Petco(2) measurement in nontracheally intubated, spontaneously breathing patients.
    Anesthesia and analgesia 04/2010; 110(4):1039-42. DOI:10.1213/ANE.0b013e3181d365fd · 3.47 Impact Factor
Show more