Article

Scatter correction for cone-beam CT in radiation therapy.

Department of Radiation Oncology, Stanford University, Stanford, California 94305, USA.
Medical Physics (Impact Factor: 3.01). 07/2009; 36(6):2258-68. DOI: 10.1118/1.3130047
Source: PubMed

ABSTRACT Cone-beam CT (CBCT) is being increasingly used in modern radiation therapy for patient setup and adaptive replanning. However, due to the large volume of x-ray illumination, scatter becomes a rather serious problem and is considered as one of the fundamental limitations of CBCT image quality. Many scatter correction algorithms have been proposed in literature, while a standard practical solution still remains elusive. In radiation therapy, the same patient is scanned repetitively during a course of treatment, a natural question to ask is whether one can obtain the scatter distribution on the first day of treatment and then use the data for scatter correction in the subsequent scans on different days. To realize this scatter removal scheme, two technical pieces must be in place: (i) A strategy to obtain the scatter distribution in on-board CBCT imaging and (ii) a method to spatially match a prior scatter distribution with the on-treatment CBCT projection data for scatter subtraction. In this work, simple solutions to the two problems are provided. A partially blocked CBCT is used to extract the scatter distribution. The x-ray beam blocker has a strip pattern, such that partial volume can still be accurately reconstructed and the whole-field scatter distribution can be estimated from the detected signals in the shadow regions using interpolation/extrapolation. In the subsequent scans, the patient transformation is determined using a rigid registration of the conventional CBCT and the prior partial CBCT. From the derived patient transformation, the measured scatter is then modified to adapt the new on-treatment patient geometry for scatter correction. The proposed method is evaluated using physical experiments on a clinical CBCT system. On the Catphan 600 phantom, the errors in Hounsfield unit (HU) in the selected regions of interest are reduced from about 350 to below 50 HU; on an anthropomorphic phantom, the error is reduced from 15.7% to 5.4%. The proposed method is attractive in applications where a high CBCT image quality is critical, for example, dose calculation in adaptive radiation therapy.

0 Bookmarks
 · 
245 Views
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Abstract Objective: To evaluate the impact of tube voltage, tube current, pulse number, and magnification factor on the image quality of a novel experimental set-up and the corresponding radiation. Materials and Methods: Six human temporal bones with cochlear implant were imaged using various tube voltages, tube currents, pulse numbers, and magnification. The effect of radiation was evaluated using a metaloxide semiconductor field-effect transistor (MOSFET) dosimeter device on an anthropomorphic RANDO RAN102 male head phantom. A copper and aluminum combination filter was used for hardware filtration. Results: Overall, 900 frames, 11 mA, and 88 kV provided the best image quality. In temporal bones imaged with the optimized parameters, the cochlea, osseous spiral lamina, modiolus, stapes, round window niche, and oval window landmarks were demonstrated with anatomic structures still fully assessable in all parts and acceptable image quality. The most dominant contributor to the effective dose was bone marrow (36%-37 %) followed by brain (34%-36%), remainder tissues (12%), extra-thoracic airways (7%), and oral mucosa (5%). Conclusions: By increasing the number of frames, the image quality of the inner ear details obtained using the novel cone-beam computed tomography improved.
    The Annals of otology, rhinology, and laryngology 12/2014; · 1.05 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: On an onboard cone-beam CT (CBCT) system, the poly-energetic beam generated by current commercial x-ray tubes hardens as it penetrates the object. The beam-hardening effect results in CT image artifacts of up to 100 HU, especially around dense objects (e.g. bones). We propose to insert a half-blocked beam filter between x-ray source and object, such that any ray passing through the object is filtered once from one of the opposite directions in a single full scan. The conventional data processing of dual energy imaging is then applied to obtain material decomposition as well as CT images with no beam-hardening artifacts. In this preliminary study, we demonstrate the method feasibility using computer simulations. The beam filter thickness is optimized to balance the CT number accuracy and the noise level on the reconstructed images via both analytical calculation and measurement of image noise. A calibration phantom is designed to obtain the decomposition function for energy selective imaging by projection measurements without the knowledge of x-ray spectrum and detector response. Using the optimized beam blocker and the calibration phantom, we simulate the data acquisition using scan settings of a commercial CBCT system. Our proposed approach significantly suppresses beam-hardening artifacts and reduces the image error from 65 HU to 9 HU in the selected regions of interest on a head phantom. The method also successfully decomposes bone and soft tissue, with 95% accuracy.
    SPIE Medical Imaging; 03/2013
  • [Show abstract] [Hide abstract]
    ABSTRACT: A beam stop array (BSA) is widely used to estimate the scatter distribution of cone-beam computed tomography (CT) experimentally. A rectangular BSA, which is the most generally used, produces a penumbra region at the edge of the grid on the projection image. The penumbra region can be the cause of an inaccurate scatter estimate and loss of image data. In this study, two types of BSAs have been designed and developed, a linear BSA (l-BSA), which is a simple rectangular type, and a curved BSA (c-BSA) to avoid the penumbra effect. A Monte Carlo simulation was performed to determine the thickness and the material of beam stop arrays. The particle tracking technique of MCNP5 code was used to analyze the accuracy of the scatter distribution acquired by the beam stop arrays. The improvement in the image quality was analyzed by using an in-house phantom, a simple type CT phantom, and a humanoid phantom. How the curved beam stop array improved the penumbra effect was also analyzed in a large field of view. The additional dose caused by a dual scan was analyzed by using a Monte Carlo simulation and experiment. The accuracies of the scatter estimates were 83.4% for the c-BSA and 90.4% for the l-BSA. The c-BSA compensated the penumbra effect effectively while the l-BSA had a severe penumbra region in the partial projection image in a large field of view. Contrasts were improved by 24.8 (air-PMMA), 55.8 (Teflon-PMMA), and 81.7% (water-PMMA) for the c-BSA and 11.1 (air-PMMA), 44.1 (Teflon-PMMA), and 82.2% (water-PMMA) for the l-BSA by scatter correction. After noise suppression, the contrast-to-noise ratios were improved by 2.7–4.1 times for the c-BSA and 3.5–5.3 times for the l-BSA. Uniformities were also improved by up to 10 times for the c-BSA and 12 times for the l-BSA. The quality of the humanoid phantom image was also improved after the scatter correction. Consequently, the c-BSA can improve the image quality of cone-beam CT and compensate for the penumbra effect without an additional dose of radiation.
    Journal- Korean Physical Society 04/2014; 64(8):1220-1229. · 0.43 Impact Factor

Full-text (2 Sources)

Download
9 Downloads
Available from
Aug 22, 2014