Article

Intraoperative use of cone-beam computed tomography in a cadaveric ossified cochlea model

Departments of Otolaryngology-Head and Neck Surgery and Surgical Oncology, Princess Margaret Hospital, University of Toronto, Toronto, Ontario, Canada.
Otolaryngology Head and Neck Surgery (Impact Factor: 1.72). 06/2009; 140(5):697-702. DOI: 10.1016/j.otohns.2008.12.046
Source: PubMed

ABSTRACT To describe a cadaveric temporal bone model of labyrinthitis ossificans and investigate the utility of intraoperative cone-beam computed tomography (CBCT) in the facilitating cochlear implantation.
Cadaveric temporal bone study.
Five cadaveric heads had cement introduced into the 10 cochleas. CBCT and a conventional CT scan were compared to assess the extent of cochlear obliteration. The cement was drilled-out (under CBCT guidance, if required) and cochlear implant electrode arrays (from 3 different manufacturers) inserted.
CBCT images demonstrated temporal bone anatomy and the extent of cochlear obliteration as clearly as conventional CT in all cases. Intraoperative CBCT guided drilling and facilitated electrode placement in two of five heads (3 of 10 ears). Streak-artifact from the electrodes of two devices partially obscured image clarity.
The obliterated cochlear model reproduced a disease-ossified cochlear both radiographically and surgically. CBCT is useful for intraoperative imaging to facilitate electrode array placement in the obliterated or congenitally abnormal cochlea.

0 Followers
 · 
144 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: A prototype system for intraoperative cone beam computed tomography (CBCT) imaging has been developed and augmented with real time optical tracking of a surgical drill. We hypothesize that this system provides sufficient accuracy for guidance of temporal bone surgery. Basic research. Measurements of drill localization accuracy using CBCT imaging were obtained with a custom three-dimensional calibration object. Integrated CBCT imaging and drill tracking were prospectively evaluated on 12 cadaver temporal bones. Six inexperienced and six experienced surgeons conducted four surgical tasks: cortical mastoidectomy, posterior tympanotomy, cochleostomy, and a translabyrinthine approach to the internal auditory canal. Questionnaires provided expert feedback on tracking accuracy and system usability. Target registration error measurements of drill tracking accuracy and precision yielded a mean of 0.76 mm, a maximum of 1.30 mm, and a standard deviation of 0.21 mm. Anatomical landmark identification tasks (e.g., facial nerve, incus, semicircular canals, cochlea) provided additional validation of system accuracy. The usability and utility of the guidance system were positively rated by both groups of surgeons, with further modifications underway to improve tracking line of sight and registration workflow. Experienced but in particular inexperienced surgeons indicated significant benefits in cases involving extensive disease, abnormal anatomy, and loss of anatomical landmarks. The integration of intraoperative CBCT imaging with optical tracking provides sufficient accuracy to localize anatomical structures within the temporal bone using an otological drill. Future studies will explore the role of this technology in complex oncological resections, in surgery for congenital anomalies, and as a tool for teaching. 4. Laryngoscope, 2013.
    The Laryngoscope 11/2013; 123(11). DOI:10.1002/lary.24130 · 2.03 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Objective Conventional computed tomography (CT) imaging is the standard imaging technique for temporal bone diseases, whereas cone-beam CT (CBCT) imaging is a very fast imaging tool with a significant less radiation dose compared with conventional CT. We hypothesize that a system for intraoperative cone-beam CT provides comparable image quality to diagnostic CT for identifying temporal bone anatomical landmarks in cadaveric specimens.Study DesignCross-sectional study.SettingUniversity tertiary care facility.Subjects and Methods Twenty cadaveric temporal bones were affixed into a head phantom and scanned with both a prototype cone-beam CT C-arm and multislice helical CT. Imaging performance was evaluated by 3 otologic surgeons and 1 head and neck radiologist. Participants were presented images in a randomized order and completed landmark identification questionnaires covering 21 structures.ResultsCBCT and multislice CT have comparable performance in identifying temporal structures. Three otologic surgeons indicated that CBCT provided statistically equivalent performance for 19 of 21 landmarks, with CBCT superior to CT for the chorda tympani and inferior for the crura of the stapes. Subgroup analysis showed that CBCT performed superiorly for temporal bone structures compared with CT. The radiologist rated CBCT and CT as statistically equivalent for 18 of 21 landmarks, with CT superior to CBCT for the crura of stapes, chorda tympani, and sigmoid sinus.ConclusionCBCT provides comparable image quality to conventional CT for temporal bone anatomical sites in cadaveric specimens. Clinical applications of low-dose CBCT imaging in surgical planning, intraoperative guidance, and postoperative assessment are promising but require further investigation.
    Otolaryngology Head and Neck Surgery 10/2013; 150(1). DOI:10.1177/0194599813510862 · 1.72 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Nonlinear partial volume (NLPV) effects can be significant for objects with large attenuation differences and fine detail structures near the spatial resolution limits of a tomographic system. This is particularly true for small metal devices like cochlear implants. While traditional model-based approaches might alleviate these artifacts through very fine sampling of the image volume and subsampling of rays to each detector element, such solutions can be extremely burdensome in terms of memory and computational requirements. The work presented in this paper leverages the model-based approach called “known-component reconstruction” (KCR) where prior knowledge of a surgical device is integrated into the estimation. In KCR, the parameterization of the object separates the volume into an unknown background anatomy and a known component with unknown registration. Thus, one can model projections of an implant at very high spatial resolution while limiting the spatial resolution of the anatomy - in effect, modeling NLPV effects where they are most significant. We present modifications of the KCR approach that can be used to largely eliminate NLPV artifacts, and demonstrate the efficacy of the modified technique (with improved image quality and accurate implant position estimates) for the cochlear implant imaging scenario.
    SPIE Medical Imaging 2013, Lake Buena Vista, Florida, USA; 03/2013