Article

The phase I multicenter trial (STAPLE-1) of the Aptus endovascular repair system: results at 6 months and 1 year.

Georgetown University Hospital, Washington, DC, USA.
Journal of vascular surgery: official publication, the Society for Vascular Surgery [and] International Society for Cardiovascular Surgery, North American Chapter (impact factor: 3.52). 05/2009; 49(4):851-7; discussion 857-8. DOI:10.1016/j.jvs.2008.10.064 pp.851-7; discussion 857-8
Source: PubMed

ABSTRACT This phase I IDE study (STAPLE-1) evaluated the primary endpoints of safety (major device-related adverse events at 30 days) and feasibility (successful deployment of all endograft components) of the Aptus Endovascular abdominal aortic aneurysm (AAA) Repair System (Aptus Endosystems, Inc, Sunnyvale, Calif) to treat AAAs.
A prospective, single arm Federal Drug Administration (FDA) Phase I IDE study was performed. The Aptus endograft is a three-piece modular device with a flexible unsupported main body and two fully supported limbs in a 5.3 mm outer diameter (OD) (16F) delivery system for all iliac limbs and two of three main body sizes. The largest main body (29 mm diameter) is in a 6 mm (18 F OD) delivery system. EndoStaples measuring 4 mm (length) by 3 mm (diameter) designed to provide transmural graft fixation to the adventitia are applied independent of the endograft delivery system. Inclusion criteria included a proximal aortic neck length of 12 mm and iliac landing zone of 10 mm. Secondary endpoints included freedom from endoleaks, rupture, migration, and device integrity.
Twenty-one (21) patients were enrolled at five centers. All patients received the Aptus Endograft and EndoStaples. Ninety-six EndoStaples (range, 2-10; median, 4) were implanted. All patients (n = 21) completed 1-month and 6-month follow-up evaluation and 14 completed 1-year follow-up. Two proximal cuffs and one limb extension were used as adjunctive endograft components at implantation. Three secondary interventions were performed in 2 patients for limb thrombosis. There were no EndoStaple-related adverse events, device integrity failures, migrations, or conversions.
These results of the STAPLE-1 trial document the acute safety and feasibility of the Aptus Endograft and EndoStaples. Early follow-up demonstrates excellent 6-month and 1-year results. A pivotal phase II trial is underway at 25 US centers.

0 0
 · 
0 Bookmarks
 · 
31 Views

Keywords

1-year results
 
6-month follow-up evaluation
 
adjunctive endograft components
 
Aptus endograft
 
device integrity
 
device integrity failures
 
endograft delivery system
 
EndoStaple-related adverse events
 
excellent 6-month
 
flexible unsupported main body
 
iliac landing zone
 
iliac limbs
 
main body sizes
 
major device-related adverse events
 
pivotal phase II trial
 
primary endpoints
 
Secondary endpoints
 
secondary interventions
 
three-piece modular device
 
transmural graft fixation