Article

A new weighting function to achieve high temporal resolution in circular cone-beam CT with shifted detectors.

Institute of Medical Physics, University of Erlangen-Nürnberg, Erlangen, Germany.
Medical Physics (impact factor: 2.83). 01/2009; 35(12):5898-909. pp.5898-909
Source: PubMed

ABSTRACT The size of the field of measurement (FOM) in computed tomography is limited by the size of the x-ray detector. In general, the detector is mounted symmetrically with respect to the rotation axis such that the transaxial FOM diameter approximately equals the lateral dimensions of the detector when being demagnified to the isocenter. To enlarge the FOM one may laterally shift the detector by up to 50% of its size. Well-known weighting functions must then be applied to the raw data prior to convolution and backprojection. In this case, a full scan or a scan with more than 360 degrees angular coverage is required to obtain complete data. However, there is a small region, the inner FOM, that is covered redundantly and where a partial scan reconstruction may be sufficient. A new weighting function is proposed that allows one to reconstruct partial scans in that inner FOM while it reconstructs full scan or overscan data for the outer FOM, which is the part that contains no redundancies. The presented shifted detector partial scan algorithm achieves a high temporal resolution in the inner FOM while maintaining truncation-free images for the outer part. The partial scan window can be arbitrarily shifted in the angular direction, what corresponds to shifting the temporal window of the data shown in the inner FOM. This feature allows for the reconstruction of dynamic CT data with high temporal resolution. The approach presented here is evaluated using simulated and measured data for a dual source micro-CT scanner with rotating gantry.

0 0
 · 
0 Bookmarks
 · 
60 Views
  • Source
    Conference Proceeding: Low-dose phase-correlated cone-beam micro-CT of small animals
    [show abstract] [hide abstract]
    ABSTRACT: Double-gated in-vivo small animal cone-beam micro-CT scans provide five-dimensional information about the object: the three volume dimensions plus the temporal dimensions of the respiratory motion and the heart motion, respectively. Double gating is typically performed to separate respiratory from cardiac motion when imaging the animal's lung or heart.
    Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE; 12/2010
  • Source
    Article: Low-dose cardio-respiratory phase-correlated cone-beam micro-CT of small animals.
    [show abstract] [hide abstract]
    ABSTRACT: Micro-CT imaging of animal hearts typically requires a double gating procedure because scans during a breath-hold are not possible due to the long scan times and the high respiratory rates, Simultaneous respiratory and cardiac gating can either be done prospectively or retrospectively. True five-dimensional information can be either retrieved with retrospective gating or with prospective gating if several prospective gates are acquired. In any case, the amount of information available to reconstruct one volume for a given respiratory and cardiac phase is orders of magnitud lower than the total amount of information acquired. For example, the reconstruction of a volume from a 10% wide respiratory and a 20% wide cardiac window uses only 2% of the data acquired. Achieving a similar image quality as a nongated scan would therefore require to increase the amount of data and thereby the dose to the animal by up to a factor of 50. To achieve the goal of low-dose phase-correlated (LDPC) imaging, the authors propose to use a highly efficient combination of slightly modified existing algorithms. In particular, the authors developed a variant of the McKinnon-Bates image reconstruction algorithm and combined it with bilateral filtering in up to five dimensions to significantly reduce image noise without impairing spatial or temporal resolution. The preliminary results indicate that the proposed LDPC reconstruction method typically reduces image noise by a factor of up to 6 (e.g., from 170 to 30 HU), while the dose values lie in a range from 60 to 500 mGy. Compared to other publications that apply 250-1800 mGy for the same task [C. T. Badea et al., "4D micro-CT of the mouse heart," Mol. Imaging 4(2), 110-116 (2005); M. Drangova et al., "Fast retrospectively gated quantitative four-dimensional (4D) cardiac micro computed tomography imaging of free-breathing mice," Invest. Radiol. 42(2), 85-94 (2007); S. H. Bartling et al., "Retrospective motion gating in small animal CT of mice and rats," Invest. Radiol. 42(10), 704-714 (2007)], the authors' LDPC approach therefore achieves a more than tenfold dose usage improvement. The LDPC reconstruction method improves phase-correlated imaging from highly undersampled data. Artifacts caused by sparse angular sampling are removed and the image noise is decreased, while spatial and temporal resolution are preserved. Thus, the administered dose per animal can be decreased allowing for long-term studies with reduced metabolic inference.
    Medical Physics 03/2011; 38(3):1416-24. · 2.83 Impact Factor
  • Source
    Conference Proceeding: Simple ROI Cone--Beam Computed Tomography
    Nuclear Science Symposium Conference Record (NSS/MIC); 12/2010

Full-text

View
2 Downloads
Available from

Keywords

360 degrees angular coverage
 
angular direction
 
backprojection
 
complete data
 
computed tomography
 
detector partial scan algorithm
 
dual source micro-CT scanner
 
dynamic CT data
 
full scan
 
inner FOM
 
lateral dimensions
 
new weighting function
 
outer FOM
 
overscan data
 
partial scan reconstruction
 
partial scan window
 
raw data
 
rotating gantry
 
transaxial FOM diameter
 
Well-known weighting functions