Taser Blunt Probe Dart-To-Heart Distance Causing Ventricular Fibrillation in Pigs

Department of Electrical and Computer Engineering, University of Wisconsin, Madison, WI 53706, USA.
IEEE transactions on bio-medical engineering (Impact Factor: 2.35). 01/2009; 55(12):2768-71. DOI: 10.1109/TBME.2008.2002154
Source: PubMed


The maximum distance between the heart and a model Taser stimulation dart, called the dart-to-heart distance, at which the Taser can directly cause ventricular fibrillation (VF), was measured in pigs. A 9-mm-long blunt probe was advanced snugly through the surrounding tissues toward the heart. Five animals [pig mass=61.2+/-6.23 standard deviation (SD) kg] for ten dart-to-heart distances where the Taser caused VF were tested. The dart-to-heart distances where the Taser caused VF of the first stimulation site ranged from 4 to 8 mm with average 6.2 mm+/-1.79 (SD) and of the second stimulation site ranged from 2 to 8 mm with average 5.4 mm+/-2.41 (SD). The results help inform the evolving discussion of risks associated with Tasers.