Article

Combined optical imaging and mammography of the healthy breast: Optical contrast derived from breast structure and compression

Massachusetts General Hospital, Charlestown, MA 02148 USA.
IEEE transactions on medical imaging 02/2009; 28(1):30-42. DOI: 10.1109/TMI.2008.925082
Source: PubMed

ABSTRACT In this paper, we report new progress in developing the instrument and software platform of a combined X-ray mammography/diffuse optical breast imaging system. Particularly, we focus on system validation using a series of balloon phantom experiments and the optical image analysis of 49 healthy patients. Using the finite-element method for forward modeling and a regularized Gauss-Newton method for parameter reconstruction, we recovered the inclusions inside the phantom and the hemoglobin images of the human breasts. An enhanced coupling coefficient estimation scheme was also incorporated to improve the accuracy and robustness of the reconstructions. The recovered average total hemoglobin concentration (HbT) and oxygen saturation (SO2) from 68 breast measurements are 16.2 microm and 71%, respectively, where the HbT presents a linear trend with breast density. The low HbT value compared to literature is likely due to the associated mammographic compression. From the spatially co-registered optical/X-ray images, we can identify the chest-wall muscle, fatty tissue, and fibroglandular regions with an average HbT of 20.1+/-6.1 microm for fibroglandular tissue, 15.4+/-5.0 microm for adipose, and 22.2+/-7.3 microm for muscle tissue. The differences between fibroglandular tissue and the corresponding adipose tissue are significant (p < 0.0001). At the same time, we recognize that the optical images are influenced, to a certain extent, by mammographical compression. The optical images from a subset of patients show composite features from both tissue structure and pressure distribution. We present mechanical simulations which further confirm this hypothesis.

Download full-text

Full-text

Available from: Dana H Brooks, Aug 31, 2015
0 Followers
 · 
149 Views
 · 
75 Downloads
  • Source
    • "To overcome these intrinsic shortcomings, multi-modal approaches that incorporate the hemoglobin-DOT with other imaging techniques have been intensively investigated, where the feasible-region information provided by the other modalities is normally used to reasonably regularize the inverse problem of the hemoglobin-DOT. Thus far, advances in multi-modal approaches using standard imaging modality of considerably high spatial resolution have been made in three fronts: concurrent optical measurements with MRI [17,18], X-ray mammography [19], and US [20]. "
    [Show abstract] [Hide abstract]
    ABSTRACT: A combined time-domain fluorescence and hemoglobin diffuse optical tomography (DOT) system and the image reconstruction methods are proposed for enhancing the reliability of breast-dedicated optical measurement. The system equipped with two pulsed laser diodes at wavelengths of 780 nm and 830 nm that are specific to the peak excitation and emission of the FDA-approved ICG agent, and works with a 4-channel time-correlated single photon counting device to acquire the time-resolved distributions of the light re-emissions at 32 boundary sites of tissues in a tandem serial-to-parallel mode. The simultaneous reconstruction of the two optical (absorption and scattering) and two fluorescent (yield and lifetime) properties are achieved with the respective featured-data algorithms based on the generalized pulse spectrum technique. The performances of the methodology are experimentally assessed on breast-mimicking phantoms for hemoglobin- and fluorescence-DOT alone, as well as for fluorescence-guided hemoglobin-DOT. The results demonstrate the efficacy of improving the accuracy of hemoglobin-DOT based on a priori fluorescence localization.
    Biomedical Optics Express 02/2013; 4(2):331-48. DOI:10.1364/BOE.4.000331 · 3.50 Impact Factor
  • Source
    • "Women still suffer from about 1 in 8 lifetime risk of developing invasive breast cancer [1]. Despite technology development in the field of mammography [1], [2] in recent years, breast cancer is still an unsolved problem. However, Mammography has been one of the most reliable methods for early detection and diagnosis of breast cancers [3]. "
    [Show abstract] [Hide abstract]
    ABSTRACT: A novel cognitive classifier has been introduced to develop a trustable mammography Computer Aided Diagnosis (CADx) system which is called Cognitive Weighted Linear Aggregation (CWLA). A group of in-depth analyzed features are extracted from the preprocessed Regions of Interest (ROIs) and mapped from set of real numbers to a set of linguistic terms. The proposed classifier primes a knowledge base which is developed according to a mammography expert. The semantic comparison of the extracted features with the expectations of the knowledge base, which is called cognitive resonance, leads to a primary clustering. Finally, the linguistic terms are remapped onto the set of real numbers and the final assessment comes out from the weighted linear aggregation of clustered categories. Since the output of the system comes with reason, the system is reliable. The achieved area under Receiver Operational Characteristics (ROC) curve (Az) and False Positive Rate (FPR) are 0.858 and 5.26%, respectively.
    Biomedical Engineering (ICBME), 2011 18th Iranian Conference of, Tehran, Iran; 12/2011
  • Source
    • "We test the proposed regularization algorithm with a set of healthy subjects measured with our combined DOT/DBT imaging system. The details of the experiments and protocols are described in [21] and [22]. Both radio-frequency modulated (RF) and continuous-wave (CW) lasers at 685 nm and 830nm were used for the data acquisition. "
    [Show abstract] [Hide abstract]
    ABSTRACT: The development of effective multi-modality imaging methods typically requires an efficient information fusion model, particularly when combining structural images with a complementary imaging modality that provides functional information. We propose a composition-based image segmentation method for X-ray digital breast tomosynthesis (DBT) and a structural-prior-guided image reconstruction for a combined DBT and diffuse optical tomography (DOT) breast imaging system. Using the 3D DBT images from 31 clinically measured healthy breasts, we create an empirical relationship between the X-ray intensities for adipose and fibroglandular tissue. We use this relationship to then segment another 58 healthy breast DBT images from 29 subjects into compositional maps of different tissue types. For each breast, we build a weighted-graph in the compositional space and construct a regularization matrix to incorporate the structural priors into a finite-element-based DOT image reconstruction. Use of the compositional priors enables us to fuse tissue anatomy into optical images with less restriction than when using a binary segmentation. This allows us to recover the image contrast captured by DOT but not by DBT. We show that it is possible to fine-tune the strength of the structural priors by changing a single regularization parameter. By estimating the optical properties for adipose and fibroglandular tissue using the proposed algorithm, we found the results are comparable or superior to those estimated with expert-segmentations, but does not involve the time-consuming manual selection of regions-of-interest.
    Biomedical Optics Express 08/2010; 1(1):223-235. DOI:10.1364/boe.1.000223 · 3.50 Impact Factor
Show more