Article

Robust microwave-assisted extraction protocol for determination of total mercury and methylmercury in fish tissues.

Department of Chemistry and Biochemistry, Duquesne University, Pittsburgh, PA 15282, USA.
Analytica chimica acta (impact factor: 4.31). 02/2009; 631(2):121-8. DOI:10.1016/j.aca.2008.10.044
Source: PubMed

ABSTRACT A rapid and efficient closed vessel microwave-assisted extraction (MAE) method based on acidic leaching was developed and optimized for the extraction of total mercury (Hg), inorganic mercury (Hg(2+)) and methylmercury (CH(3)Hg(+)) from fish tissues. The quantitative extraction of total Hg and mercury species from biological samples was achieved by using 5molL(-1) HCl and 0.25molL(-1) NaCl during 10min at 60 degrees C. Total Hg content was determined using inductively coupled plasma mass spectrometry (ICP-MS). Mercury species were measured by liquid chromatography hyphenated with inductively coupled plasma mass spectrometry (LC-ICP-MS). The method was validated using biological certified reference materials ERM-CE464, DOLT-3, and NIST SRM-1946. The analytical results were in good agreement with the certified reference values of total Hg and CH(3)Hg(+) at a 95% confidence level. Further, accuracy validation using speciated isotope-dilution mass spectrometry (SIDMS, as described in the EPA Method 6800) was carried out. SIDMS was also applied to study and correct for unwanted species transformation reactions during and/or after sample preparation steps. For the studied reference materials, no statistically significant transformation between mercury species was observed during the extraction and determination procedures. The proposed method was successfully applied to fish tissues with good agreement between SIDMS results and external calibration (EC) results. Interspecies transformations in fish tissues were slightly higher than certified reference materials due to differences in matrix composition. Depending on the type of fish tissue, up to 10.24% of Hg(2+) was methylated and up to 1.75% of CH(3)Hg(+) was demethylated to Hg(2+).

0 0
 · 
0 Bookmarks
 · 
42 Views
  • Article: Trace-Level Automated Mercury Speciation Analysis.
    [show abstract] [hide abstract]
    ABSTRACT: An automated system for methyl Hg analysis by purge and trap gas chromatography (GC) was evaluated, with comparison of several different instrument configurations including chromatography columns (packed column or capillary), detector (atomic fluorescence, AFS, or inductively coupled plasma mass spectrometry, ICP-MS, using quadrupole and sector field ICP- MS instruments). Method detection limits (MDL) of 0.042 pg and 0.030 pg for CH(3)Hg(+) were achieved with the automated Hg analysis system configured with AFS and ICPMS detection, respectively. Capillary GC with temperature programming was effective in improving resolution and decreasing retention times of heavier Hg species (in this case C(3)H(7)Hg(+)) although carryover between samples was increased. With capillary GC, the MDL for CH(3)Hg(+) was 0.25 pg for AFS detection and 0.060 pg for ICP-MS detection. The automated system was demonstrated to have high throughput (72 samples analyzed in 8 hours) requiring considerably less analyst time than the manual method for methyl mercury analysis described in EPA 1630.
    Analytical methods 01/2011; 3(5):1143-1148. · 1.55 Impact Factor

Keywords

95% confidence level
 
analytical results
 
biological certified reference materials ERM-CE464
 
biological samples
 
certified reference materials
 
certified reference values
 
inorganic mercury
 
Interspecies transformations
 
liquid chromatography hyphenated
 
matrix composition
 
Mercury species
 
plasma mass spectrometry
 
quantitative extraction
 
sample preparation steps
 
SIDMS results
 
speciated isotope-dilution mass spectrometry
 
statistically significant transformation
 
total mercury
 
unwanted species transformation reactions
 
vessel microwave-assisted extraction