Osseointegration of zirconia implants compared with titanium: an
in vivo study
Rita Depprich1, Holger Zipprich2, Michelle Ommerborn*3,
Christian Naujoks†1, Hans-Peter Wiesmann4, Sirichai Kiattavorncharoen5,
Hans-Christoph Lauer2, Ulrich Meyer1, Norbert R Kübler†1 and
Jörg Handschel1
Address: 1Department of Cranio- and Maxillofacial Surgery, Heinrich-Heine-University, Düsseldorf, Germany, 2Department of Prosthetic
Dentistry, Section of Materials Sciences, Johann Wolfgang Goethe University, Frankfurt, Germany, 3Department of Operative and Preventive
Dentistry and Endodontics, Heinrich-Heine-University, Düsseldorf, Germany, 4Department of Cranio- and Maxillofacial Surgery, Westfalian
Wilhelms-University, Münster, Germany and 5Department of Oral and Maxillo-Facial Surgery, Mahidol University, Bangkok, Thailand
Email: Rita Depprich - depprich@med.uni-duesseldorf.de; Holger Zipprich - zipprich@em.uni-frankfurt.de;
Michelle Ommerborn* - ommerborn@med.uni-duesseldorf.de; Christian Naujoks - christian.naujoks@med.uni-duesseldorf.de; Hans-
Peter Wiesmann - HansPeter.Wiesmann@ukmuenster.de; Sirichai Kiattavorncharoen - kiattav@hotmail.com; Hans-
Christoph Lauer - H.C.Lauer@em.uni-frankfurt.de; Ulrich Meyer - ulrich.meyer@med.uni-duesseldorf.de ;
Norbert R Kübler - kuebler@med.uni-duesseldorf.de; Jörg Handschel - handschel@med.uni-duesseldorf.de
* Corresponding author †Equal contributors
Abstract
Background: Titanium and titanium alloys are widely used for fabrication of dental implants. Since
the material composition and the surface topography of a biomaterial play a fundamental role in
osseointegration, various chemical and physical surface modifications have been developed to
improve osseous healing. Zirconia-based implants were introduced into dental implantology as an
altenative to titanium implants. Zirconia seems to be a suitable implant material because of its
tooth-like colour, its mechanical properties and its biocompatibility. As the osseointegration of
zirconia implants has not been extensively investigated, the aim of this study was to compare the
osseous healing of zirconia implants with titanium implants which have a roughened surface but
otherwise similar implant geometries.
Methods: Forty-eight zirconia and titanium implants were introduced into the tibia of 12 minipigs.
After 1, 4 or 12 weeks, animals were sacrificed and specimens containing the implants were
examined in terms of histological and ultrastructural techniques.
Results: Histological results showed direct bone contact on the zirconia and titanium surfaces.
Bone implant contact as measured by histomorphometry was slightly better on titanium than on
zirconia surfaces. However, a statistically significant difference between the two groups was not
observed.
Conclusion: The results demonstrated that zirconia implants with modified surfaces result in an
osseointegration which is comparable with that of titanium implants.
Published: 11 December 2008
Head & Face Medicine 2008, 4:30 doi:10.1186/1746-160X-4-30
Received: 8 July 2008
Accepted: 11 December 2008
This article is available from: http://www.head-face-med.com/content/4/1/30
© 2008 Depprich et al; licensee BioMed Central Ltd.
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0),
which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.Page 1 of 8
(page number not for citation purposes)
Page 2
Head & Face Medicine 2008, 4:30 http://www.head-face-med.com/content/4/1/30
Background
Since their introduction over 40 years ago, dental
implants have become an established treatment modality
that had revolutionized the concept of replacing missing
teeth. The recent material of choice for manufacturing
dental implants is commercially pure titanium, because of
its excellent biocompatibilty and mechanical properties
[1]. However, the gray colour of the titanium may be dis-
advantageous and give rise to esthetic problems, espe-
cially if the soft tissue situation is not optimal and the
dark colour shines through the thin periimplant mucosa
[2].
The success of endosseous implants is directly related to
the principle of osseointegration, a process of implant-
bone interaction that finally leads to bone-to-implant
anchorage [3]. As the surface topography of a biomaterial
has a major impact on osseointegration, various chemical
and physical surface modifications have been developed
to improve osseous healing of implants. Increased surface
roughness of dental implants resulted in greater bone
apposition [4] and reduced healing time [5].
Zirconia ceramics (yttrium-stabilized tetragonal poly-
crystals) seem to be a suitable material for dental implants
because of their tooth-like colour, their excellent mechan-
ical properties and their good biocompatibility [6]. They
have extensively been used as ball heads in total hip
replacements with remarkable clinical outcomes [7].
Recent animal studies have also shown successful bone
healing of dental zirconia implants under both unloaded
and loaded conditions [2,8-10]. As the conventional fab-
rication of zirconia rods usually results in realtively
smooth surfaces, only few studies have investigated rough
surface modifications of zirconia implants. This is a criti-
cal aspect, since it has been already demonstrated that sur-
face roughness and topography also influence
osseointegration of zirconia implants [6,11-13].
In comparison with titanium implants, much less is
known about the role played by surface modifications on
the osseointegration of zirconia implants. Thus, the aim
of the present study was to examine the osseous healing of
zirconia implants with acid-etched surface structures in
comparison with titanium implants.
Methods
Experimental animals
Twelve minipigs (> 5 years, average body weight 66.5 kg)
were used in this study. The investigation was approved by
the Animal Ethics Committee at the University of Düssel-
dorf, Germany. The animals were kept in small groups in
purpose-designed sties and fed on a standard diet. Twelve
Implant system
Twenty-four screw-type zirconia implants (yttrium-stabi-
lized tetragonal poly-crystals) with modified (acid-
etched) surfaces (Ra = 0.598 μm, according to manufac-
turer) were used and compared to twenty-four implants
made of commercially pure titanium with acid-etched sur-
faces (Ra = 1.77 μm, according to manufacturer).
Implants were supplied by Konus Dental Implants (Bin-
gen, Germany). All implants had the same macroscopic
design with a standardized diameter of 3.5 mm and a
length of 9 mm.
Surgical procedure
All surgeries were performed under sterile conditions in a
veterinary operating theatre. The animals were sedated by
an intramuscular injection (10 mg/kg) of ketamine (Ket-
avet®, Pfizer, Karlsruhe, Germany), 1 ml atropine (Atro-
pinsulfat Braun®, Braun, Melsungen, Germany) and 5 mg/
kg azaperone (Stresnil®, Janssen-Cilag, Neuss, Germany).
Anesthesia was induced with an intravenous bolus of 3–5
ml thiopental (Thiopental inresa®, Inresa Arzneimittel,
Freiburg, Germany) followed by intubation and mainte-
nance of anaesthesia by inhalation of 1.5% isoflurane. For
analgesia animals received 0.5 ml piritramide (Dipido-
lor®, Janssen-Cilag, Neuss, Germany). In the areas to be
exposed to surgery, 5 ml of local anaesthesia [articain
hydrochloride, (Ultracain® DS, 1:200.000), Aventis,
Frankfurt, Germany] was injected. The tibias were exposed
by skin incisions and via fascial-periosteal flaps. Thereaf-
ter, four implants were placed in the tibia. The implant
sites were sequentially enlarged with two drills according
to the standard protocol of the manufacturer. Implants
measuring 9 mm in length and 3.5 mm in diameter were
inserted using continuous external sterile saline irrigation
to minimize bone damage caused by overheating. At the
surgical site, the skin and the fascia-periosteum were
closed in separate layers with single resorbable sutures
(Vicryl®2-0, Ethicon, Norderstedt, Germany). Periopera-
tively, the animals received amoxicillin (10 mg/kg KG)
(Duphamox LA®, Fort Dodge, Würselen, Germany) as
antibiotic and carprofen p.o. (4.4 mg/kg) (Rimadyl®,
Pfizer, Karlsruhe, Germany) as antiphlogistic medication
for three days. The animals were inspected after the first
few postoperative days for signs of wound dehiscence or
infection and, thereafter, weekly to assess general health.
After 1, 4 or 12 weeks animals were sacrificed (4 minipigs
each) with an overdose of pentobarbital (Eutha 77®, Essex
Pharma, München, Germany) given intravenously. Fol-
lowing euthanasia, tibia block specimens containing the
implants and surrounding tissues were dissected from the
animals. The block samples were sectioned with a saw to
remove unnecessary fragments of bone and soft tissue andPage 2 of 8
(page number not for citation purposes)
hours before surgery animals were denied feed although
water was accessible ad libitum.
were prepared for the subsequent investigations.
Page 3
Head & Face Medicine 2008, 4:30 http://www.head-face-med.com/content/4/1/30
Histological analyses
The implants were immediately fixed in 4% buffered for-
maldehyde for approximately one week. Then the speci-
mens were dehydrated in a graded series of ethanol.
Thereafter, samples were embedded in methyl metacrylate
(Technovit®7200, Heraeus Kulzer, Dormagen, Germany).
With the help of the cutting-grinding technique according
to Donath, longitudinal sections were ground to about
20–40 μm for conventional microscopy (Exakt Apparate-
bau, Norderstedt, Germany). Two central histological sec-
tions of each implant were obtained and samples were
stained with toluidine blue and Masson-Trichrome-Gold-
ner. The slides were examined and photographed with a
Leica DM 5000B (Leica Microsystems, Wetzlar, Germany)
light microscope, equipped with a Leica DC 300F high
resolution camera.
Histomorphometry
Histomorphometric evaluation was performed after one
central slice was chosen at 50-fold magnification using a
digital camera. The software ImageJ 1.37v® (open source:
http://rsb.info.nih.gov/ij/features.html) was used to
measure the bone-to-implant contact (BIC) ratio, defined
as the length of bone surface border in direct contact with
the implant (× 100 (%)).
Statistical analysis
All calculations were performed with the help of SPSS for
Windows (SPSS Inc., Chicago, IL, USA). The results from
the histomorphometric measurements were expressed as
means ± standard deviations. The different treatment
groups were compared using a Mann-Whitney U test. A p
< 0.05 was set for significance.
Results
The animals recovered well after surgery and no signs of
infection were noted upon clinical examination at any
time during the observation period (Figure 1). Light
microscopical analysis demonstrated that matrix-rich
regeneration tissue displaced the blood clot between the
implant surface and the bone tissue in the first week after
surgical procedure (Figure 2). After 4 weeks, mature regen-
eration tissue with formation of osteoid and woven bone
was observed (Figure 3). Close contact of the bone to the
implant was seen both on titanium and zirconia surfaces.
Circumferential bone tissue formation was detectable on
the zirconia implant surface. After 12 weeks of healing,
hard tissue integration of the titanium as well as the zirco-
nia implants was achieved. Mature lamellar bone in direct
contact to the titanium and zirconia implants was found
(Figure 4). No signs of inflammation were detected in any
of the specimens. Histologically detectable minor differ-
ences between the zirconia and the titanium implants
were no longer evident.
The bone-to-implant contact increased over the examina-
tion period for both zirconia and titanium implants (Fig-
ure 5). After 1 week of healing, the mean BIC was 35.3%
± 10.8 for the zirconia and 47.7% ± 9.1 for the titanium
implants, respectively. After 4 weeks in situ, BIC of the zir-
conia implants averaged 45.3% ± 15.7 and 58.6% ± 9.5
for the titanium implants. After 12 weeks the BIC values
Radiograph showing titanium (left) and zirconia (right) implants inserted into the tibia of minipigs after 12 weeks of healing timeFigure 1
Radiograph showing titanium (left) and zirconia (right) implants inserted into the tibia of minipigs after 12 Page 3 of 8
(page number not for citation purposes)
weeks of healing time.
Page 4
Head & Face Medicine 2008, 4:30 http://www.head-face-med.com/content/4/1/30
were 71.4% ± 17.8 for the zirconia implants and 82.9% ±
10.7 for the titanium implants. There were no statistically
significant differences observed betweeen the titanium
and zirconia implants (p < 0.05) in regards to bone-to-
implant contact after 1, 4 or 12 weeks.
Discussion
Zirconia is a bioinert nonresorbable metal oxide that
offers mechanical properties which are superior over other
ceramic biomaterials, e.g. high fracture toughness and
bending strength [14]. Because of its good chemical and
material stability, high strength and resilience it seems to
be a suitable material for dental application [7]. Its suc-
cessful application in dentistry for fabricating endodontic
posts and for crown and bridge restorations has been
reported in several studies [15-17]. Especially because of
its tooth-like colour, zirconia was suggested to be a desir-
able alternative material to titanium for the fabrication of
dental implants. The results of the present study have
shown that zirconia implants fabricated with a modified
surface seem to be integrated into bone in a similar fash-
ion as titanium.
After one week of healing, distinct gaps between the
implant and the bone filled with matrix-rich regeneration
tissue were observed in a few locations. After 4 weeks,
Micrograph showing matrix-rich regeneration tissue (orange) between the implant and bone (green)Figure 2
Micrograph showing matrix-rich regeneration tissue (orange) between the implant and bone (green). Zirconia
implant (left), titanium implant (right) (Masson-Trichrome-Goldner, 100-fold).
zirconia titanium
At 4 weeks after implantation, osteoid and woven bone were formed both on zirconia (left) and titanium implant surfaces (right) (toluidine blue, 50-fold)Fi ure 3
At 4 weeks after implantation, osteoid and woven bone were formed both on zirconia (left) and titanium
zirconia titaniumPage 4 of 8
(page number not for citation purposes)
implant surfaces (right) (toluidine blue, 50-fold).
Page 5
Head & Face Medicine 2008, 4:30 http://www.head-face-med.com/content/4/1/30
woven bone, and after 12 weeks, lamellar bone, was visi-
ble in intimate contact with the implant surfaces. A loose
connective tissue layer separating bone tissue and the zir-
conia surface as described by Sennerby et al. [6] previ-
ously, was not found in our samples.
Osseointegration of threaded zirconia implants has been
recently investigated by Rothamel et al. [18]. They com-
pared the osseous healing of zirconia implants with mod-
ified (machined and sand blasted) implant surfaces from
polished commercially pure titanium. After 4 days of
healing time, a distinct gap between bone tissue and the
implant surface filled with remodelling blood clot was
noticed. Two weeks after implantation, woven bone grow-
ing in the direction of the implant was observed, followed
by the formation of lamellar bone after 28 days. When the
healing period was nearly completed after 8 weeks, inti-
mate contact of lamellar bone to the implant surface was
evident. However, the barrier resulting from the original
gap was still visible with many osteoblasts bridging the
gap, which indicates a high biocompatibility of the used
implant materials.
The results of the present study also showed an increasing
BIC over the healing period. However, there were no sta-
tistically significant histomorphometrical differences
observed between zirconia and titanium implants. This
finding is in accordance with other animal studies which
also failed to demonstrate differences between structured
zirconia and titanium implant surfaces [2,6,8,12,19],
likely secondary to the fact that zirconia is highly biocom-
several authors [2,6,10,18]. The reported differences in
BIC seem to be attributable to different animal models
(dogs, monkeys, rabbits and minipigs) used for the exper-
iments [2,9,20]. In order to establish standardized condi-
tions for the histomorphometric analysis, implants were
placed in the tibia since this bone has constant bone
geometries over a longer distance. Therefore, the BIC only
depends on the implant osseointegration and not on the
bone features at the implantation site. In contrast to the
results from a similar study [21], there were no detach-
ment or separation of bone tissue and the zirconia surface
with loose connective tissue detectable at any time.
The BIC measured in our study (45.3% after 4 weeks)
showed similar results as demonstrated by Sennerby et al.
[6]. The authors demonstrated a BIC of 36% for the non-
modified zirconia implants and BICs of more than 45%
for the zirconia implants with surface modification after 6
weeks of healing in the tibia of rabbits.
Scarano et al. [10] observed 68% BIC of the untreated zir-
conia implants after 4 weeks in the tibia of rabbits. After 6
months of unloaded healing in the mandibles of dogs,
Dubruille et al. [9] measured a BIC of 65% for the zirconia
implants compared with 68% of alumina implants and
54% of the titanium implants. The surface topography of
the implants in these studies was not investigated. Kohal
and coworkers [2] determined slightly higher BIC values
after implant insertion into the maxillae of monkeys fol-
lowed by 5 months of loaded healing (68% for sand-
blasted zirconia implants and 73% for sandblasted and
After 12 weeks of healing, mature lamellar bone is evident in intimate contact with the zirconia implant (left) and titanium implant (right) (toluidine blue, 100-fold)Figure 4
After 12 weeks of healing, mature lamellar bone is evident in intimate contact with the zirconia implant (left)
and titanium implant (right) (toluidine blue, 100-fold).
zirconia titaniumPage 5 of 8
(page number not for citation purposes)
patible. An average BIC > 60%, which was achieved after
4 weeks following implantation, had been reported by
acid-etched titanium implants). However, the surface
topography was not measured or described. In the present
End of preview.