Article

Wear of a 5 Megarad Cross-linked Polyethylene Liner: A 6-year RSA Study

Discipline of Orthopaedics and Trauma, University of Adelaide, Adelaide, SA, Australia, .
Clinical Orthopaedics and Related Research (Impact Factor: 2.88). 01/2013; 471(7). DOI: 10.1007/s11999-013-2789-x
Source: PubMed

ABSTRACT BACKGROUND: One cross-linked polyethylene (XLPE) liner is manufactured using a lower dose of radiation, 5 Mrad, which may result in less cross-linking. The reported in vivo wear rate of this XLPE liner in patients undergoing THA has varied, and has included some patients in each reported cohort who had greater than 0.1 mm/year of wear, which is an historical threshold for osteolysis. Previous studies have measured wear on plain radiographs, an approach that has limited sensitivity. QUESTIONS/PURPOSES: We therefore measured the amount and direction of wear at 6 years using Radiostereometric analysis (RSA) in patients who had THAs that included a cross-linked polyethylene liner manufactured using 5 Mrad radiation. METHODS: We prospectively reviewed wear in 30 patients who underwent primary THAs with the same design of cross-linked acetabular liner and a 28-mm articulation. Tantalum markers were inserted during surgery and all patients had RSA radiographic examinations at 1 week, 6 months, 1, 2, and 6 years postoperatively. RESULTS: The mean proximal, two-dimensional (2-D) and three-dimensional (3-D) wear rates calculated between 1 year and 6 years were 0.014, 0.014, and 0.018 mm/per year, respectively. The direction of the head penetration recorded between 1 week and 6 years was in a proximal direction for all patients, proximolateral for 16 of 24 patients, and proximomedial for eight of 24 patients. CONCLUSIONS: The proximal, 2-D and 3-D wear of a XLPE liner produced using 5 Mrad of radiation was low but measurable by RSA after 6 years. No patients had proximal 2-D or 3-D wear rates exceeding 0.1 mm/year. Further followup is needed to evaluate the effect of XLPE wear particles on the development of long-term osteolysis.

Download full-text

Full-text

Available from: Stuart A Callary, Nov 20, 2014
0 Followers
 · 
81 Views
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: A sequentially irradiated and annealed, second-generation highly crosslinked polyethylene (XLPE) liner was introduced clinically in 2005 to reduce in vivo oxidation. This liner design has also been shown to reduce wear in vitro when compared with conventional and first-generation crosslinked liners. To date, there is only one study reporting an in vivo wear rate of this liner at 5 years' followup. However, that study used measurements made from plain radiographs, which have limited sensitivity, particularly when monitoring very low amounts of wear. What is the amount and direction of wear at 5 years using radiostereometric analysis (RSA) in patients who had THAs that included second-generation XLPE? We prospectively reviewed 21 patients who underwent primary cementless THA with the same design of XLPE acetabular liner and 32-mm articulation. Tantalum markers were inserted during surgery and all patients had RSA radiographs at 1 week, 6 months, and 1, 2, and 5 years postoperatively. Femoral head penetration within the acetabular component was measured with UmRSA(®) software. One patient died and two had incomplete radiographs leaving 18 radiographic series for analysis. The mean amounts of proximal, two-dimensional, and three-dimensional head penetration between 1 week and 5 years were 0.018, 0.071, and 0.149 mm, respectively. The mean proximal, two-dimensional, and three-dimensional wear rates calculated between 1 year and 5 years were all less than 0.001 mm/year with no patient recording a wear rate of more than 0.040 mm/year. The head penetration of a second-generation XLPE liner remained low at 5 years and the wear rate calculated after the first year was low in all directions. This low level of wear remains encouraging for the future clinical performance of this material. Level IV, therapeutic study. See Instructions for Authors for a complete description of levels of evidence.
    Clinical Orthopaedics and Related Research 07/2013; 471(11). DOI:10.1007/s11999-013-3188-z · 2.88 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Background and purpose — Wear rates of highly crosslinked polyethylene (XLPE) acetabular components have varied considerably between different published studies. This variation is in part due to the different techniques used to measure wear and to the errors inherent in measuring the relatively low amounts of wear in XLPE bearings. We undertook a scoping review of studies that have examined the in vivo wear of XLPE acetabular components using the most sensitive method available, radiostereometric analysis (RSA). Methods — A systematic search of the PubMed, Scopus, and Cochrane databases was performed to identify published studies in which RSA was used to measure wear of XLPE components in primary total hip arthroplasty (THA). Results — 18 publications examined 12 primary THA cohorts, comprising only 260 THAs at 2–10 years of follow-up. The mean or median proximal wear rate reported ranged from 0.00 to 0.06 mm/year. However, differences in the manner in which wear was determined made it difficult to compare some studies. Furthermore, differences in RSA methodology between studies, such as the use of supine or standing radiographs and the use of beaded or unbeaded reference segments, may limit future meta-analyses examining the effect of patient and implant variables on wear rates. Interpretation — This scoping review confirmed the low wear rates of XLPE in THA, as measured by RSA. We make recommendations to enhance the standardization of reporting of RSA wear results, which will facilitate early identification of poorly performing implants and enable a better understanding of the effects of surgical and patient factors on wear.
    Acta Orthopaedica 10/2014; DOI:10.3109/17453674.2014.972890 · 2.45 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: We hypothesized that a vitamin E-treated polyethylene articulating liner had lower femoral head penetration than a conventional liner. Fifty-one patients with primary hip osteoarthritis were randomized to an uncemented cup with either vitamin E-treated highly cross-linked polyethylene (HXLPE) (vitamin-e group, n = 25) or a standard HXLPE liner (control group, n = 26) and followed for two years after surgery. The outcome variable was polyethylene (PE) linear head penetration, measured with radiostereometry (RSA). The head penetration in x- (medial/lateral) and y- (vertical) axes was significantly lower for the vitamin E-treated liner up to two years with a mean difference between the groups of 0.10 mm (95 % CI 0.04-0.18) and 0.08 mm (95 % CI 0.01-0.15), respectively. The initial total penetration was lower for the vitamin-E group but did not reach statistical significance at two years with a mean difference between the groups of 0.07 mm (95 % CI -0.001 to 0.15, p = 0.09). We found no differences in clinical outcome between the groups. In this prospective, randomized, controlled trial on a new vitamin-E diffused HXLPE liner we found, compared to a conventional HXLPE liner, lower initial head penetration and lower superior and medial wear. A longer follow-up than two years is necessary to demonstrate lower overall head penetration in vitamin-E liners compared to contemporary highly cross-linked PE.
    International Orthopaedics 01/2015; DOI:10.1007/s00264-015-2680-3 · 2.02 Impact Factor