Development of a dual modality imaging system: A combined gamma camera and optical imager

Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, 50 Ilwon-Dong, Gangnam-Gu, Seoul 135-710, Korea.
Physics in Medicine and Biology (Impact Factor: 2.76). 06/2009; 54(14):4547. DOI: 10.1088/0031-9155/54/14/011


Several groups have reported the development of dual modality Gamma camera/optical imagers, which are useful tools for investigating biological processes in experimental animals. While previously reported dual modality imaging instrumentation usually employed a separated gamma camera and optical imager, we designed a detector using a position sensitive photomultiplier tube (PSPMT) that is capable of imaging both gamma rays and optical photons for combined gamma camera and optical imager. The proposed system consists of a parallel-hole collimator, an array-type crystal and a PSPMT. The top surface of the collimator and array crystals is left open to allow optical photons to reach the PSPMT. Pulse height spectra and planar images were obtained using a Tc-99m source and a green LED to estimate gamma and optical imaging performances. When both gamma rays and optical photon signals were detected, the signal interferences caused by each other signal were evaluated. A mouse phantom and an ICR mouse containing a gamma ray and optical photon source were imaged to assess the imaging capabilities of the system. The sensitivity, energy resolution and spatial resolution of the gamma image acquired using Tc-99m were 1.1 cps/kBq, 26% and 2.1 mm, respectively. The spatial resolution of the optical image acquired with an LED was 3.5 mm. Signal-to-signal interference due to the optical photon signal in the gamma pulse height spectrum was negligible. However, the pulse height spectrum of the optical photon signal was found to be affected by the gamma signal, and was obtained between signals generated by gamma rays with a correction using a veto gate. Gamma ray and optical photon images of the mouse phantom and ICR mouse were successfully obtained using the single detector. The experimental results indicated that both optical photon and gamma ray imaging are feasible using a detector based on the proposed PSPMT.

17 Reads
  • Source
    • "In the present study, we used a simple approach by moving a mouse support between different instruments. However, other groups are attempting to build an integrated apparatus that can perform at the same time both optical and nuclear imaging and at the same time using rings of separate scintigraphy detectors and CCD cameras [34], combined PET or gamma camera with an optical detector [35, 36] or adaptation of a conical mirror-based fDOT inside the PET gantry [37, 38]. Such instrument could improve fDOT/PET imaging and will be useful to perform dynamic imaging. "
    [Show abstract] [Hide abstract]
    ABSTRACT: Purpose Given the progress of fluorescence diffuse optical tomography (fDOT) technology, here, we study the additional benefits provided by multimodal PET/fDOT imaging by comparing the biodistribution of 2-deoxy-2-[18F]fluoro-d-glucose ([18F]FDG) in tumors with three fluorescent probes: a glucose analog, a protease activatable optical probe, and a ligand of αvβ3 integrin. Procedures Sequential fDOT/PET/computed tomography (CT) imaging of mice was performed with a custom multimodal mouse support that allows the subject to be transferred between the fDOT and the PET/CT scanners. Experiments were performed in xenografted tumor models derived from the human breast cancer line MDA-MB 231 and compared to ex vivo analysis. Results The three-dimensional signals showed that the fluorescent glucose analog is not colocalized with [18F]FDG, raising questions about its use as a surrogate probe of the PET tracer. Fusion of [18F]FDG with the other fluorescent probes showed evidence of high variability both for the protease activity and the αvβ3 integrin expression during tumor growth. Conclusion The added value of hybrid PET/fDOT over the two modalities was demonstrated for cross-validation of probes and for better characterization of tumor models.
    Molecular imaging and biology: MIB: the official publication of the Academy of Molecular Imaging 08/2012; 15(3). DOI:10.1007/s11307-012-0581-z · 2.77 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: A new positron emission tomography (PET) detector signal processing method, the charge signal transmission approach, is proposed for the development of a hybrid PET-magnetic resonance imaging (MRI). A number of experiments were performed to demonstrate that the Geiger-mode avalanche photodiode (GAPD) charge output could be transmitted to a preamplifier using a long cable without degrading the PET signal performance. A PET module consisted of LYSO and a GAPD with a 4 x 4 array. The GAPD output was transmitted to the preamplifier through flexible flat cables. The effect of the cable length on the PET performance was examined using seven different lengths ranging from 10 to 300 cm outside and inside the 7 T animal MRI. Four parameters (rise time, fall time, amplitude, and area of the preamplifier output) were measured as a function of the cable length using a 10 GS/s oscilloscope and three parameters (photopeak position, energy resolution, and time resolution) were measured using a 100 MS/s DAQ unit. The effect of the cable length on the MR phantom images was investigated. In addition, the effect of the PET module configuration on its temperature stability was assessed by acquiring the energy and time spectra. There were no significant changes in the PET module performance as a function of the cable length, both outside and inside MRI. The performance changes in energy information, such as the amplitude, area, photopeak position, and energy resolution, were <3% with cable lengths ranging from 10 to 300 cm and the change in the time resolution was <6%. There were no obvious artifacts or changes in the line profile in the MR phantom images. Moreover, no manifest changes in the photopeak position and coincidence counting rate were observed in the PET modules employing the charge signal transmission approach, whereas considerable degradation of the PET module performance was observed in the voltage signal transmission approach. This study demonstrated that it is feasible to design a hybrid PET-MRI using the charge signal transmission approach, which is expected to have more advantages than other approaches.
    Medical Physics 11/2010; 37(11):5655-64. DOI:10.1118/1.3495683 · 2.64 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: A dual-modality imaging system for simultaneous fluorescence molecular tomography (FMT) and positron emission tomography (PET) of small animals has been developed. The system consists of a noncontact 360°-projection FMT module and a flat panel detector pair based PET module, which are mounted orthogonally for the sake of eliminating cross interference. The FMT images and PET data are simultaneously acquired by employing dynamic sampling mode. Phantom experiments, in which the localization and range of radioactive and fluorescence probes are exactly indicated, have been carried out to verify the feasibility of the system. An experimental tumor-bearing mouse is also scanned using the dual-modality simultaneous imaging system, the preliminary fluorescence tomographic images and PET images demonstrate the in vivo performance of the presented dual-modality system.
    IEEE Transactions on Nuclear Science 03/2011; 58(1-58):51 - 57. DOI:10.1109/TNS.2010.2068310 · 1.28 Impact Factor
Show more


17 Reads
Available from