Article

Prosthetic misfit of implant-supported prosthesis obtained by an alternative section method.

Department of Dental Materials and Prosthodontics, Dental School of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil.
The journal of advanced prosthodontics 05/2012; 4(2):89-92. DOI:10.4047/jap.2012.4.2.89 pp.89-92
Source: PubMed

ABSTRACT Adequate passive-fitting of one-piece cast 3-element implant-supported frameworks is hard to achieve. This short communication aims to present an alternative method for section of one-piece cast frameworks and for casting implant-supported frameworks.
Three-unit implant-supported nickel-chromium (Ni-Cr) frameworks were tested for vertical misfit (n = 6). The frameworks were cast as one-piece (Group A) and later transversally sectioned through a diagonal axis (Group B) and compared to frameworks that were cast diagonally separated (Group C). All separated frameworks were laser welded. Only one side of the frameworks was screwed.
The results on the tightened side were significantly lower in Group C (6.43 ± 3.24 µm) when compared to Groups A (16.50 ± 7.55 µm) and B (16.27 ± 1.71 µm) (P<.05). On the opposite side, the diagonal section of the one-piece castings for laser welding showed significant improvement in the levels of misfit of the frameworks (Group A, 58.66±14.30 µm; Group B, 39.48±12.03 µm; Group C, 23.13±8.24 µm) (P<.05).
Casting diagonally sectioned frameworks lowers the misfit levels. Lower misfit levels for the frameworks can be achieved by diagonally sectioning one-piece frameworks.

0 0
 · 
0 Bookmarks
 · 
55 Views
  • Article: The significance of passive framework fit in implant prosthodontics: current status.
    [show abstract] [hide abstract]
    ABSTRACT: The clinical and laboratory procedures employed for framework fabrication are inadequate to provide an absolute passive fit for implant-supported fixed superstructures. Although some prosthetic complications are attributed to the lack of passive fit, its effect on implant success is questionable. Nevertheless, the clinical results of increasing applications of advanced technology to improve framework fit seem promising. This article reviews the clinical significance of passive fit and the factors that affect the final fit of implant-supported frameworks.
    Implant Dentistry 02/2001; 10(2):85-92. · 1.05 Impact Factor
  • Article: Marginal accuracy of four-unit zirconia fixed dental prostheses fabricated using different computer-aided design/computer-aided manufacturing systems.
    [show abstract] [hide abstract]
    ABSTRACT: Besides load-bearing capacity, marginal accuracy is a further crucial factor influencing the clinical long-term reliability of fixed dental prostheses (FDPs). The aim of this in vitro study was to evaluate the marginal fit of four-unit zirconia bridge frameworks fabricated using four different computer-aided design (CAD)/computer-aided manufacturing (CAM) systems. Ten frameworks were manufactured using each fabricating system. Three systems (inLab, Everest, Cercon) processed white-stage zirconia blanks, which had to be sintered to final density after milling, while with one system (Digident) restorations were directly milled from a fully sintered material. After manufacturing, horizontal and vertical marginal discrepancies, as well as the absolute marginal discrepancy, were determined by means of a replica technique. The absolute marginal discrepancy, which is considered to be the most suitable parameter reflecting restorations' misfit in the marginal area, had a mean value of 58 mum for the Digident system. By contrast, mean absolute marginal discrepancies for the three other systems, processing presintered blanks, differed significantly and ranged between 183 and 206 mum. Within the limitations of this study, it could be concluded that the marginal fit of zirconia FDPs is significantly dependent on the CAD/CAM system used, with restorations processed of fully sintered zirconia showing better fitting accuracy.
    European Journal Of Oral Sciences 07/2009; 117(3):319-25. · 1.88 Impact Factor
  • Article: Failures and complications in 391 consecutively inserted fixed prostheses supported by Brånemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup.
    [show abstract] [hide abstract]
    ABSTRACT: A total of 391 edentulous maxillae and mandibles was consecutively treated with routine fixed prostheses, supported by 2,199 implants. The patients were followed for 1 year; the overall success rate was 99.5% and 98.1% for the prostheses and implants, respectively. The number of severe complications was low, but it was possible to identify a significantly higher ratio of problems in the maxillae throughout the observation period than for the mandibles. While many problems occurred in both the maxillae and mandibles, there were also problems more typical for each jaw. Diction and fractures of resin teeth were more common problems in maxillae; cheek and lip biting was a more frequent postinsertion complication in the treatment of mandibles. Most problems were easily resolved, and the retrievability of the prostheses was of great advantage.
    The International journal of oral & maxillofacial implants 02/1991; 6(3):270-6. · 1.78 Impact Factor

Full-text (2 Sources)

View
4 Downloads
Available from
8 Nov 2012

Keywords

3-element implant-supported frameworks
 
Adequate passive-fitting
 
alternative method
 
Casting diagonally sectioned frameworks
 
diagonal axis
 
diagonal section
 
diagonally
 
diagonally sectioning one-piece frameworks
 
Group B
 
Group C
 
implant-supported frameworks
 
laser welding
 
Lower misfit levels
 
misfit levels
 
one-piece
 
one-piece castings
 
short communication
 
Three-unit implant-supported nickel-chromium
 
transversally sectioned
 
vertical misfit