Article

Evaluation of entropy for monitoring the depth of anesthesia compared with bispectral index: a multicenter clinical trial.

Department of Anesthesiology, Beijing Chaoyang Hospital, Capital Medical University, Beijing 100020, China.
Chinese medical journal (impact factor: 0.86). 04/2012; 125(8):1389-92. pp.1389-92
Source: PubMed

ABSTRACT As a new electroencephalogram (EEG) signal processing technique for monitoring the depth of anesthesia, entropy consists of two indices: reaction entropy (RE) and state entropy (SE). Our study compared entropy with classical bispectral index (BIS) in reduction of myoelectrical interference and noxious stimuli with EEG signals.
Two hundred and eighty patients (ASA I-II, 18-60 years old) undergoing scheduled surgeries from seven medical centers were enrolled. Anesthesia induction was managed with propofol via the target-controlled infusion (TCI) system. The results of BIS, RE, SE, mean arterial pressure (MAP) and heart rate (HR) were recorded before anesthesia induction, at the moment of unconsciousness, before and 2 minutes after administration of muscle relaxant, and before and one and three minutes after the tracheal intubation.
The values of half maximum effective concentrations (EC50), 5% effective concentrations (EC05) and 95% effective concentrations (EC95) of propofol effect-site concentration at the onset of unconsciousness were 1.2 (1.1-1.3 µg/ml), 2.5 (2.4-2.5 µg/ml) and 3.7 (3.7-3.8 µg/ml), while those of the predicted plasma propofol concentration were 2.8 (2.7-2.9 µg/ml), 3.9 (3.8-3.9 µg/ml) and 4.9 (4.8-5.0 µg/ml), respectively. The values of BIS, SE and RE were 62, 59 and 63 when 50% of patients lost consciousness, and 79, 80, 85 and 42, 37, 44, respectively, when 5% and 95% of patients were unconscious. The values of BIS, RE and SE dropped two minutes after the injection of muscle relaxant, but there were no significant differences between RE and SE. MAP and HR increased visibly, which indicated a reaction to tracheal intubation; the values of BIS, RE and SE, however, did not display any significant changes.
This large-sample multicentric study confirmed the values of RE and SE as approximating BIS value, at the onset of unconsciousness during propofol TCI anesthesia. After elimination of myoelectrical activation, all values of RE, SE and BIS decreased significantly and the three indices were less sensitive to noxious stimuli than cardiovascular responses.

0 0
 · 
0 Bookmarks
 · 
36 Views

Keywords

2 minutes
 
5% effective concentrations
 
95% effective concentrations
 
Anesthesia induction
 
approximating BIS value
 
ASA I-II
 
classical bispectral index
 
large-sample multicentric study
 
maximum effective concentrations
 
muscle relaxant
 
myoelectrical interference
 
new electroencephalogram
 
predicted plasma propofol concentration
 
propofol effect-site concentration
 
propofol TCI anesthesia
 
reaction entropy
 
significant changes
 
state entropy
 
three indices
 
tracheal intubation
 

Jian-Dong Gao