Nano-flow multidimensional liquid chromatography platform integrated with combination of protein and peptide separation for proteome analysis.

Key Laboratory of Separation Science for Analytical Chemistry, National Chromatographic R&A Center, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian, P R China.
Journal of Separation Science (Impact Factor: 2.59). 05/2012; 35(14):1764-70. DOI: 10.1002/jssc.201200052
Source: PubMed

ABSTRACT An integrated multidimensional nano-flow liquid chromatography platform with the combination of protein and peptide separation via online digestion by an immobilized enzymatic reactor was established, and successfully applied for proteome analysis. By this platform, proteins were first separated by a weak anion and weak cation mixed-bed microcolumn under a series of salt steps, online digested by a trypsin immobilized enzymatic reactor, digests trapped and desalted by a C18 precolumn, separated by nano-reversed phase liquid chromatography, and finally identified by electrospray ionization-MS/MS. To evaluate the performance of such a platform, Escherichia coli whole cell lysate proteins were analyzed. Compared with the results obtained by shotgun approach, the identified protein number was increased by 6%, with the consumed time decreased from 38 to 14 h. We also compared with integrate platform based on micro-HPLC, and the required sample amount was decreased to 8 μg. These results demonstrated that such an integrated approach would be an attractive alternative to commonly applied approaches for proteome research.

0 Bookmarks
 · 
127 Views
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: In this paper, we present a method for the determination of low- and ultra-low copy-number proteins in biomaterials based on a combination of concentrating the protein from the sample onto CNBr-activated Sepharose 4B (via non-specific binding of free amino groups) and multiple reaction monitoring (MRM). The detection limit and the dependence of the MRM peak areas on the concentration of protein in the sample were determined using the proteins CYP 102 and BSA, as a model system, both in solution and after their addition to human plasma. Non-specific protein enrichment of proteins from diluted sample volumes of 10 to 50 mL was found to increase the range of linear dependence of the chromatographic peak area on concentration by more than three orders of magnitude, allowing a lower limit of detection limit (LLOD) of as low as 10(-18) M. At this LLOD, at least two tryptic peptides of CYP102 and BSA could be detected with signal-to-noise ratios of ≥7.0. The results were equally good for samples containing pure protein mixtures and proteins spiked into diluted depleted human blood plasma.
    Proteomics 12/2012; · 3.97 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: The E. coli proteome was digested with trypsin and fractionated using solid phase extraction on a C18 SPE column. Seven fractions were collected and analyzed by capillary zone electrophoresis (CZE)-electrospray ionization tandem mass spectrometry (ESI-MS/MS). The separation was performed in a 60 cm long linear polyacrylamide-coated capillary with a 0.1% (v/v) formic acid separation buffer. An electrokinetic sheath-flow electrospray interface was used to couple the separation capillary with an Orbitrap Velos operating in higher-energy collisional dissociation mode. Each CZE-ESI-MS/MS run lasted 50 minutes and total MS time was 350 minutes. A total of 23,706 peptide spectra matches, 4,902 peptide IDs, and 871 protein group IDs were generated using MASCOT with false discovery rate less than 1% on the peptide level. The total mass spectrometer analysis time was less than six hours, the sample identification rate (145 proteins/hour) was more than two times higher than previous studies of the E. coli proteome, and the amount of sample consumed (<1 μg) was roughly four-fold less than previous studies. These results demonstrate that CZE is a useful tool for the bottom-up analysis of prokaryote proteomes. This article is protected by copyright. All rights reserved.
    Proteomics 06/2013; · 3.97 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: In this work, a novel kind of organic–silica hybrid monolith based immobilized enzymatic reactor (IMER) was developed. The monolithic support was prepared by a single step “one-pot” strategy via the polycondensation of tetramethoxysilane and vinyltrimethoxysilane and in situ copolymerization of methacrylic acid and vinyl group on the precondensed siloxanes with ammonium persulfate as the thermal initiator. Subsequently, the monolith was activated by N-(3-dimethylaminopropyl) – N′-ethylcarbodiimide (EDC) and N-hydroxysuccinimide (NHS), followed by the modification of branched polyethylenimine (PEI) to improve the hydrophilicity. Finally, after activated by EDC and NHS, trypsin was covalently immobilized onto the monolithic support. The performance of such a microreactor was evaluated by the in sequence digestion of bovine serum albumin (BSA) and myoglobin, followed by MALDI-TOF-MS analysis. Compared to those obtained by traditional in-solution digestion, not only higher sequence coverages for BSA (74 ± 1.4% vs. 59.5 ± 2.7%, n = 6) and myoglobin (93 ± 3% vs. 81 ± 4.5%, n = 6) were obtained, but also the digestion time was shortened from 24 h to 2.5 min, demonstrating the high digestion efficiency of such an IMER. The carry-over of these two proteins on the IMER was investigated, and peptides from BSA could not be found in mass spectrum of myoglobin digests, attributed to the good hydrophilicity of our developed monolithic support. Moreover, the dynamic concentration range for protein digestion was proved to be four orders of magnitude, and the IMER could endure at least 7-day consecutive usage. Furthermore, such an IMER was coupled with nano-RPLC–ESI/MS/MS for the analysis of extracted proteins from Escherichia coli. Compared to formerly reported silica hybrid monolith based IMER and the traditional in-solution counterpart, by our developed IMER, although the identified protein number was similar, the identified distinct peptide number was improved by 7% and 25% respectively, beneficial to improve the reliability of protein identification. The IMER was further online integrated with two-dimensional nano-HPLC–MS/MS system for the analysis of protein extracts from hepatocellular carcinoma (HCC) cells with low metastasis rate, and more than 3000 protein groups were identified, with only 46 proteins identified from the residues of the IMER. All these results demonstrated that such a hybrid monolith based IMER would be of great promise in the high throughput and high confidence proteome analysis.
    Journal of Chromatography A 10/2014; · 4.26 Impact Factor