Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease.

Shandong Provincial Key Laboratory of Diagnosis and Treatment of Cardio-cerebral Vascular Diseases, Shandong Medical Imaging Research Institute, Shandong University, No.324, Jingwu Road, Jinan, Shandong, 250021, China.
European Radiology (impact factor: 3.22). 05/2012; 22(10):2057-66. DOI:10.1007/s00330-012-2479-1 pp.2057-66
Source: PubMed

ABSTRACT To compare accuracy, image quality and radiation dose between high-pitch spiral and sequential modes on 128-slice dual-source computed tomographic (DSCT) angiography in children with congenital heart disease (CHD).
Forty patients suspected with CHD underwent 128-slice DSCT angiography with high-pitch mode and sequential mode respectively. All the anomalies were confirmed by the surgical and/or the conventional cardiac angiography (CCA) findings. The diagnostic accuracy, the subjective and objective image quality and effective radiation doses were compared.
There was no significant difference in diagnostic accuracy (χ ( 2 ) = 0.963, P > 0.05), the objective parameters for image quality (P > 0.05) and the image quality of great vessels (u = 167.500, P > 0.05) between the two groups. The image quality of intracardiac structures and coronary arteries was significantly better in the sequential mode group than that in the high-pitch group (u = 112.500 and 100.000, P < 0.05). The mean effective dose in high-pitch group (0.17 ± 0.05 mSv) was significantly lower (t = 5.287, P < 0.05) than that in the sequential mode group (0.29 ± 0.09 mSv).
Both the high-pitch and the sequential modes for 128-slice DSCT angiography provide high accuracy for the assessment of CHD in children, while the high-pitch mode, even with some image quality decrease, further significantly lowers the radiation dose.
• Modern CT provides excellent anatomical detail of congenital heart disease. • Dual source CT systems offer high-pitch spiral and sequential modes. • The high-pitch mode provides high accuracy for the assessment of CHD. • A few images using the high-pitch mode were occasionally slightly degraded. • But the high-pitch mode significantly lowers the radiation dose.

0 0
 · 
0 Bookmarks
 · 
18 Views

Keywords

128-slice DSCT angiography
 
128-slice dual-source computed tomographic
 
congenital heart disease
 
conventional cardiac angiography
 
coronary arteries
 
diagnostic accuracy
 
effective radiation doses
 
excellent anatomical detail
 
great vessels
 
high-pitch mode
 
image quality decrease
 
images
 
intracardiac structures
 
objective image quality
 
sequential mode
 
sequential mode group
 
sequential modes
 
two groups
 
• Dual source CT systems offer high-pitch spiral
 
• Modern CT