Conference Paper

Can large-area avalanche photodiodes be used for a clinical PETIMRI block detector

Department of Radiology and Molecular Imaging Program, Stanford, University, CA, USA
DOI: 10.1109/NSSMIC.2008.4774466 Conference: Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Source: IEEE Xplore


We are investigating a high resolution, Angerlogic based PET block detector using large-area avalanche photodiodes (LAAPD). Such block detectors will be used in a simultaneous PET/MRI clinical scanner. Using a Monte-Carlo simulation tool, we systemically optimized the detector design taking into account the following factors: the dimensions of the scintillation crystal and the optical diffuser, the surface finish of crystals, the layout of APD arrays, and the signal-to-noise (SNR) of the system. Based on the simulation results, a block detector prototype was built with an 8×8 LYSO crystal array (crystal pitch size: 2.75 × 3.00 × 20 mm3) coupled to four APDs. The performance of the block detector with regard to the crystal resolving ability and the energy resolution is reported.

Download full-text


Available from: Hao Peng,
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Positron emission tomography (PET) is used in the clinic and in vivo small animal research to study molecular processes associated with diseases such as cancer, heart disease, and neurological disorders, and to guide the discovery and development of new treatments. This paper reviews current challenges of advancing PET technology and some of newly developed PET detectors and systems. The paper focuses on four aspects of PET instrumentation: high photon detection sensitivity; improved spatial resolution; depth-of-interaction (DOI) resolution and time-of-flight (TOF). Improved system geometry, novel non-scintillator based detectors, and tapered scintillation crystal arrays are able to enhance the photon detection sensitivity of a PET system. Several challenges for achieving high resolution with standard scintillator-based PET detectors are discussed. Novel detectors with 3-D positioning capability have great potential to be deployed in PET for achieving spatial resolution better than 1 mm, such as cadmium-zinc-telluride (CZT) and position-sensitive avalanche photodiodes (PSAPDs). DOI capability enables a PET system to mitigate parallax error and achieve uniform spatial resolution across the field-of-view (FOV). Six common DOI designs, as well as advantages and limitations of each design, are discussed. The availability of fast scintillation crystals such as LaBr(3), and the silicon photomultiplier (SiPM) greatly advances TOF-PET development. Recent instrumentation and initial results of clinical trials are briefly presented. If successful, these technology advances, together with new probe molecules, will substantially enhance the molecular sensitivity of PET and thus increase its role in preclinical and clinical research as well as evaluating and managing disease in the clinic.
    Current pharmaceutical biotechnology 05/2010; 11(6):555-71. DOI:10.2174/138920110792246555 · 2.51 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: We investigated the feasibility of designing an Anger-logic PET detector module using large-area high-gain avalanche photodiodes (APDs) for a brain-dedicated PET/MRI system. Using Monte Carlo simulations, we systematically optimized the detector design with regard to the scintillation crystal, optical diffuser, surface treatment, layout of large-area APDs, and signal-to-noise ratio (SNR, defined as the 511 keV photopeak position divided by the standard deviation of noise floor in an energy spectrum) of the APD devices. A detector prototype was built comprising an 8 × 8 array of 2.75 × 3.00 × 20.0 mm3 LYSO (lutetium-yttrium-oxyorthosilicate) crystals and a 22.0 × 24.0 × 9.0 mm3 optical diffuser. From the four designs of the optical diffuser tested, two designs employing a slotted diffuser are able to resolve all 64 crystals within the block with good uniformity and peak-to-valley ratio. Good agreement was found between the simulation and experimental results. For the detector employing a slotted optical diffuser, the energy resolution of the global energy spectrum after normalization is 13.4 ± 0.4%. The energy resolution of individual crystals varies between 11.3 ± 0.3% and 17.3 ± 0.4%. The time resolution varies between 4.85 ± 0.04 (center crystal), 5.17 ± 0.06 (edge crystal), and 5.18 ± 0.07 ns (corner crystal). The generalized framework proposed in this work helps to guide the design of detector modules for selected PET system configurations, including scaling the design down to a preclinical PET system, scaling up to a whole-body clinical scanner, as well as replacing APDs with other novel photodetectors that have higher gain or SNR such as silicon photomultipliers.
    Physics in Medicine and Biology 06/2011; 56(12):3603-27. DOI:10.1088/0031-9155/56/12/010 · 2.76 Impact Factor