A method for characterizing stapes prostheses by their mechanical transfer function.

Chair of Sensor Technology, University Erlangen-Nuremberg, Germany.
Medical Engineering & Physics (impact factor: 1.62). 03/2012; 34(5):659-63. DOI:10.1016/j.medengphy.2012.02.001
Source: PubMed

ABSTRACT In this contribution, we present and evaluate a method for characterizing stapes prostheses by their mechanical transfer function. The measurements were carried out after a stapedotomy surgery was performed in three human temporal bones conserved in 4% formaldehyde. The inner ear was drained of fluid. Successively, one of three different stapes prostheses was inserted. After such preparation, the prosthesis piston movement compared to the incus movement is measured with laser vibrometry. The magnitude transfer function considered is defined as the amplitude of the prosthesis piston movement compared to the amplitude of the incus movement. Measurements were made in a frequency range from 500Hz to 4kHz. The measured amplitudes roughly ranged between 10nm and 100nm. A great advantage of the presented method is the fact that only a small portion of the ossicular chain influences the measurement, mainly the joint between the prosthesis and the incus. Furthermore, the usage of cadaver temporal bones allows for an automated measurement setup, long term experiments and the access of measurement positions inapproachable during in vivo measurements. With this method, the different kinds of prostheses could be evaluated on incuses of different diameters.

0 0
 · 
0 Bookmarks
 · 
24 Views

Keywords

automated measurement setup
 
cadaver temporal bones
 
characterizing stapes prostheses
 
different kinds
 
frequency range
 
great advantage
 
human temporal bones conserved
 
incus movement
 
inner ear
 
magnitude transfer function
 
measured amplitudes
 
measurement positions inapproachable
 
Measurements
 
mechanical transfer function
 
presented method
 
prosthesis piston movement
 
small portion
 
stapedotomy surgery
 
term experiments
 
vivo measurements