Skull Bone Flap Fixation - Reliability and Efficacy of a New Grip-Like Titanium Device (Skull Grip) versus Traditional Sutures: A Clinical Randomized Trial

Department of Neurosurgery, Service de Neurochirurgie, Hopital Lariboisière, 75475 Paris Cedex 10, France.
min - Minimally Invasive Neurosurgery (Impact Factor: 0.62). 10/2011; 54(5-6):282-5. DOI: 10.1055/s-0031-1297246
Source: PubMed

ABSTRACT After completing a craniotomy, it is important to replace the removed bone flap in its natural position in order to guarantee brain protection as well as improve cosmesis. A skull defect can expose the brain to accidental damage, and in cases of larger defects it may also cause the patients psychosocial problems. The ideal fixation device should provide reliable attachment of the flap to the skull and promote fast bony healing to avoid possible pseudo-arthrosis and/or osteolytic changes.
This is a pilot randomized clinical trial on a series of 16 patients undergoing different craniotomies for benign brain lesions in which the bone flaps were replaced using traditional sutures (Prolene 0.0) in 8 cases and with a new skull fixation device (Skull Grip) in the other 8 (randomly allocated). All patients underwent CT scans of the head with 3D reconstruction at day 1 and day 90 postoperatively to evaluate bone flap position and fusion. These scans were independently reviewed by a neuroradiologist. Cosmesis was also evaluated clinically by the surgeon and radiologically by the neuroradiologist in the 2 patient groups.
The new "Skull Grip" device has shown stronger fixation qualities with optimal bone flap fusion and increased cosmetic healing features vs. traditional sutures.
The "Skull Grip" has shown to be a reliable, effective and stronger bone flap fixation device when compared to traditional sutures.

0 Bookmarks
 · 
108 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: After completing a craniotomy, whenever possible, it is crucial to replace and fix the removed bone flap to the cranium; this in order to keep the brain's protection as well as for cosmetic purposes. Visible skull defects might cause patients psychosocial problems and, most importantly, expose the brain to accidental damage. A fixation device should not only provide optimal attachment of the flap to the skull but also allow fast bony healing to avoid possible pseudoarthrosis and/or osteolytic changes. After performing 12 different craniotomies on 4 human cadaver heads the skull flaps were replaced using traditional sutures and a new skull fixation device; for each fixation technique a load-bearing test was performed and the results compared. Bone flaps fixed with the "Skull Grip" showed a strong fixation with optimal plastic deformation when compared to flaps held by sutures that showed less resistance to pressure and could be easily dislocated. The "Skull Grip" has shown to be a reliable, effective, and stronger bone flap fixation superior to suturing technique.
    min - Minimally Invasive Neurosurgery 05/2009; 52(2):98-100. DOI:10.1055/s-0028-1128116 · 0.62 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Craniotomy bone flaps are replaced for both cosmetic and protective purposes. The purpose of this study was to determine the facility, strength, timing, and cost effectiveness of an alternative system for securing bone flaps. The system consists of titanium discs that clamp the bone flap to the skull edge. Ten cadaver craniotomy flaps each were reattached with either 24-gauge wire, miniplates, or titanium clamps. The titanium clamp system required significantly less time to fix than either wire or miniplates. The clamps were stronger than wire and cost less than miniplates. The titanium clamp system is a reasonable alternative to present craniotomy fixation methods with respect to ease of use, time consumption, strength, and relative cost.
    Surgical Neurology 05/2000; 53(4):391-4; discussion 394-5. DOI:10.1016/S0090-3019(00)00186-5 · 1.67 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: The MRI and CT compatibility of a new design of metallic fixation implants for cranial bone flaps was investigated experimentally and following clinical application in craniotomy patients. Is imaging quality impaired through implant-induced artifacts? Possible induced movement and heating of the clamp implant as well as the magnitude of susceptibility artifacts were assessed experimentally by MRI at 1.5 T. Additionally, implant-induced artifacts were evaluated under clinical conditions for all relevant MRI-sequences in 100 craniotomy patients whose skull flaps were re-affixed using the new clamp implant. In T1-weighted spin echo and fast spin echo sequences the sizes of implant-induced artifacts were equal to the size of the implant or slightly larger; in T2-weighted gradient echo images artifacts were maximally double the size of the implant itself. Induced movement or heating was not observed. The safety in radiological imaging of titanium clamp implants for fixation of craniotomy bone flaps was established experimentally as well as under clinical conditions. The clamps can be easily identified by radiological imaging, cause no hazards to the patients, and have little effect on image quality.
    RöFo - Fortschritte auf dem Gebiet der R 01/1999; 169(6):601-4. DOI:10.1055/s-2007-1015348 · 1.96 Impact Factor