Pathologic assessment of complications with asymmetric or sulcus fixation of square-edged hydrophobic acrylic intraocular lenses.

University of Utah, Salt Lake City, UT, USA.
Ophthalmology (Impact Factor: 5.56). 03/2012; 119(5):907-13. DOI: 10.1016/j.ophtha.2011.10.022
Source: PubMed

ABSTRACT One-piece hydrophobic acrylic intraocular lenses (IOLs) are not indicated for sulcus fixation because of complications, such as uveitis-glaucoma-hyphema syndrome. Similar complications may also be observed with 3-piece lenses that have a square optic edge on the anterior surface. The objective of this study was to provide pathologic evidence of complications related to out-of-the-bag fixation of 1- or 3-piece hydrophobic acrylic IOLs with anterior and posterior square optic edges.
Comparative case series with pathology.
A total of 661 pseudophakic cadaver eyes obtained from eye banks within the United States, implanted with different IOLs.
Anterior segment scanning of whole eyes with a high-frequency ultrasound system or high-resolution anterior segment magnetic resonance imaging, followed by gross examination. Selected eyes were processed for complete histopathologic analysis; some of them were explanted before histopathology to allow for direct light microscopic evaluation of the lenses.
Findings from imaging, gross, and histopathologic evaluation that could be related to out-of-the-bag fixation of the lenses.
A total of 256 eyes were implanted with hydrophobic acrylic IOLs with anterior and posterior square optic edges; 18 of them exhibiting asymmetric or sulcus IOL fixation (six 1-piece and twelve 3-piece IOLs) underwent complete histopathologic evaluation and were compared with the contralateral eyes with symmetric in-the-bag IOL implantation. Pathologic findings were composed of IOL decentration and tilt, pigmentary dispersion within the anterior segment and on the IOL surface, iris transillumination defects, iris changes including vacuolization/disruption/loss of the pigmented layer, iris thinning, and iris atrophy, as well as synechiae and loop erosion in the case of 3-piece lenses. Findings were more significant in comparison with the control contralateral eyes and were particularly evident in relation to the sulcus-fixated haptic in the case of 1-piece lenses. The majority of the eyes with 3-piece lenses showed signs of complicated surgery.
This study provides pathologic correlation of complications that were clinically reported in the peer-reviewed literature in relation to sulcus fixation of 1-piece hydrophobic acrylic IOLs. The eyes with 3-piece lenses generally exhibited evidence of complicated surgery; therefore, all pathologic findings in those cases may not be strictly attributed to the out-of-the-bag fixation.

0 Bookmarks
 · 
94 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: To evaluate the usefulness of ultrasound biomicroscopy in confirming intraocular lens haptic-induced ocular irritation and in the management of these patients. A retrospective review of patient data. Twenty pseudophakic patients who underwent ultrasound biomicroscopy examination between May 2009 and February 2011 to confirm the clinical suspicion of misplacement of intraocular lens haptics were reviewed. Ophthalmic findings at the time of presentation and at each follow-up visit, and management of each patient were recorded. Intraocular lens haptics misplacement was confirmed by ultrasound biomicroscopy in all suspected cases. In 75% of the eyes, one haptic was embedded in the iris, in 35% it extended into the ciliary body process, and into the pars plana in 10%. Focal iris thinning/atrophy was detected by ultrasound biomicroscopy in 15% of cases, and focal angle closure in 25%. Intraocular lens exchange was performed in 40% of patients. The remaining 60% were kept under observation, with the addition of topical steroids and/or cycloplegics in eyes that demonstrated anterior chamber inflammation, and intraocular pressure lowering medications in eyes with persistent elevated intraocular pressure or glaucoma. Ultrasound biomicroscopy appears to be a valuable tool in confirming the presence of haptic-induced ocular irritation and in assisting the management of these patients.
    American Journal of Ophthalmology 01/2014; · 4.02 Impact Factor
  • Retina (Philadelphia, Pa.) 09/2014; 34(9):e25. · 2.93 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: OBJECTIVE: To evaluate the pathologic findings of 3-piece intraocular lenses (IOLs) with asymmetric or sulcus fixation in pseudophakic cadaver eyes, comparing IOLs with square or round edges on the anterior optic surface. DESIGN: Comparative case series with pathology. PARTICIPANTS: A total of 661 pseudophakic cadaver eyes, obtained from eye banks within the United States, implanted with different IOLs. METHODS: Anterior segment scanning of whole eyes with a high-frequency ultrasound system or high-resolution anterior segment magnetic resonance imaging followed by gross examination. Selected eyes were processed for complete histopathologic analysis. MAIN OUTCOME MEASURES: Findings from imaging, gross, and histopathologic evaluation that could be related to out-of-the-bag fixation of the lenses. RESULTS: Of 661 pseudophakic cadaver eyes obtained, 13 had 3-piece hydrophobic acrylic IOLs with anterior and posterior square optic edges, and 14 had 3-piece lenses with anterior round edges (13 silicone lenses and 1 hydrophobic acrylic lens) without symmetric in-the-bag fixation. These 27 selected eyes were processed for complete histopathologic analysis. Gross findings in both groups were composed of IOL decentration and tilt, pigmentary dispersion within the anterior segment and on the IOL surface, and iris transillumination defects. Histopathology of the 14 eyes with 3-piece IOLs with round anterior optic edges showed mild focal disruption of the iris pigmented layer and loop protrusion/erosion in the ciliary sulcus. Additional changes observed in the 13 eyes with square anterior optic edge IOLs included iris changes, such as vacuolization, disruption and loss of the pigmented epithelial layers, iris thinning and atrophy, synechiae, and pigmentary dispersion within the trabecular meshwork. One eye also exhibited initial signs of optic nerve disc cupping. CONCLUSIONS: In this series, pathologic findings were more severe in eyes implanted with 3-piece IOLs with square anterior optic edges, suggesting that IOLs with round anterior edges are more suitable for sulcus fixation. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found after the references.
    Ophthalmology 04/2013; · 5.56 Impact Factor