Comparison of the performance of four microtube column systems in the detection of red cell alloantibodies

Department of Transfusion Medicine and Hemostaseology, Friedrich-Alexander University Erlangen-Nürnberg, Erlangen, Germany.
Transfusion (Impact Factor: 3.57). 11/1999; 39(10):1045-50. DOI: 10.1046/j.1537-2995.1999.39101045.x
Source: PubMed

ABSTRACT The purpose of this study was to compare the performance of four currently available microtube column agglutination systems in the detection of red cell alloantibodies to that of the standard tube low-ionic-strength solution (LISS) indirect antiglobulin test (IAT) (tube LISS-IAT).
In a comparative study, 172 sera, previously demonstrated to contain red cell alloantibodies, were tested in parallel by the tube LISS-IAT and three microtube column agglutination techniques (DiaMed-ID, Ortho BioVue, and Sanofi-Pasteur Scangel) and one affinity-adherence test system (Gamma ReACT). Tests were performed simultaneously by a single person on freshly thawed sera that had been frozen at -20 degrees C.
The rate of detection of clinically significant alloantibodies (n = 154) in microtube column systems was very similar. One hundred forty-one sera (91.6%) reacted in the DiaMed-ID, 139 (90.3%) in the ReACT, 139 (90.3%) in the BioVue, and 142 (92.2%) in the Scangel. Only 117 (76.0%) of these sera reacted in the tube LISS-IAT. The detection rates for 18 antibodies of minor clinical significance (anti-M, -N, -P1, -Le(a), and -Le(b)) varied among the test systems: DiaMed-ID, 5 (28%); ReACT, 7 (39%); BioVue, 14 (78%); Scangel, 10 (56%); and tube LISS-IAT, 6 (33%). Antibody reactivity as determined by titer and score was very similar in all microtube column systems and higher in these systems than in the tube LISS-IAT.
The sensitivity of all four microtube column systems in the detection of clinically significant red cell alloantibodies was similar and was markedly superior to that of the tube LISS-IAT. An individual cost-benefit analysis should be performed in every institution to decide whether a microtube column system should be applied. If so, the antibody screen in the microtube column agglutination system should ideally be performed in advance of the crossmatch to provide time to screen for compatible units.

5 Followers
 · 
212 Views
  • Source
    • "Our parallel comparison of seven ab screening methods showed a higher sensitivity of the two solidphase systems and the four microtube column systems over the tube LISS-IAT in the detection of cs-ab with 20-min incubation at 37 °C. This is of special interest as it had been supposed previously that the sensitivity of the tube LISS-IAT might be comparable to that of microcolumn test systems if the 37 °C incubation period is extended from 10 to 20 min (Voak, 1999; Weisbach et al., 1999). Overall, these results confirm previous results obtained from investigations of much smaller numbers of antibodies (Carter et al., 1997; Issitt et al., 1997; Reilly et al., 1997). "
    [Show abstract] [Hide abstract]
    ABSTRACT: To compare the performance of seven currently available test systems in the detection of erythrocyte alloantibodies (ab), we tested in parallel 446 sera samples containing red cell ab [368 sera samples with ab that are assumed to be clinically significant (cs-ab) and 78 sera samples with ab that are assumed to be of minor clinical significance (ms-ab)] using the tube spin low-ionic-strength solution (addition method) indirect antiglobulin test (tube LISS-IAT), three microtube column agglutination techniques (DiaMed-ID, Ortho BioVue and Bio-Rad Scangel), one affinity adherence test system (CLB/Mast CellBind Screen) and two solid-phase tests [Biotest Solidscreen II and Immucor Capture-R Ready-Screen (4)]. To address the specificity of the three test systems under routine conditions, results of 4566 patient samples obtained using the tube LISS-IAT, results of 5205 patient samples obtained using the Scangel and results of 3560 samples obtained using the Capture-R were evaluated. The DiaMed-ID detected 344 cs-ab and 43 ms-ab, BioVue 333 cs-ab and 48 ms-ab, Scangel 348 cs-ab and 62 ms-ab, CellBind Screen 346 cs-ab and 47 ms-ab, Solidscreen 330 cs-ab and 38 ms-ab, Capture-R 358 cs-ab and 45 ms-ab and LISS-IAT 159 cs-ab and 12 ms-ab. In routine practice, erythrocyte cs-ab could be identified in 61 (67.8%) of 90 reactive sera (specificity: 98.6%) in the tube LISS-IAT, in 169 (58.7%) of 288 (94.4%) in Bio-Rad Scangel and in 101 (51.0%) of 198 reactive sera (94.3%) in Capture-R. We conclude that the sensitivity of the microcolumn, affinity adherence and solid-phase test systems in the detection of cs-ab was similar and was markedly superior to that of the conventional tube LISS-IAT. All high-sensitive test systems produced higher rates of false positives and ms-ab compared to the tube test. An individual cost-benefit analysis, considering the recent knowledge about the clinical significance of weak-reactive cs-ab, should be performed in every institution to decide whether and if so which high-sensitive screening system should be applied.
    Transfusion Medicine 09/2006; 16(4):276-84. DOI:10.1111/j.1365-3148.2006.00674.x · 1.31 Impact Factor
  • Source
    • "This previous information led us to evaluate the gel microcolumn assay for anti-D titration. Our data clearly indicate that the gel microcolumn assay is more sensitive than the tube technique for anti-D titration, as was expected, since the gel has been shown to be more sensitive in the detection of red cell alloantibodies (Weisbach et al., 1999). Ciavarella, Pate & Sorenson (2002) have also demonstrated that titration studies performed by gel assay detected lower known concentrations of anti-D (5 ng/ml) than traditional tube methods (11 ng/ml). "
    [Show abstract] [Hide abstract]
    ABSTRACT: Anti-D titration is the first step in the evaluation of the RhD-sensitized patient. Traditionally, anti-D titration has been performed by tube agglutination. Gel microcolumn assay is a method that has gained widespread usage throughout the world, mainly for ABO/Rh typing, unexpected antibody screening and direct antiglobulin tests. As gel assay has become widely used as a routine method to detect red blood cell alloantibodies, a critical anti-D titer needs to be established. Seventy-nine known blood samples with anti-D (titers 1-32) were titrated simultaneously by the conventional tube test and the gel microcolumn assay. Red blood cells (R0r phenotype) were used, with a final concentration of 3% for tube and 0.8% for gel. Serial twofold dilutions (2-2.048) were prepared for each technique, followed by reading in antiglobulin phase. Anti-D titration in the gel microcolumn assay showed significantly higher titers (mean 3.4-fold) than the conventional tube test in all samples studied. Based on these data, it was not possible to determine a critical titer for anti-D titration by the gel microcolumn assay.
    Clinical & Laboratory Haematology 11/2003; 25(5):311-5. DOI:10.1046/j.1365-2257.2003.00540.x · 1.30 Impact Factor
  • Source
    • "This previous information led us to evaluate the gel microcolumn assay for anti-D titration. Our data clearly indicate that the gel microcolumn assay is more sensitive than the tube technique for anti-D titration, as was expected, since the gel has been shown to be more sensitive in the detection of red cell alloantibodies (Weisbach et al., 1999). Ciavarella, Pate & Sorenson (2002) have also demonstrated that titration studies performed by gel assay detected lower known concentrations of anti-D (5 ng/ml) than traditional tube methods (11 ng/ml). "
    [Show abstract] [Hide abstract]
    ABSTRACT: There are several methods for antibody detection and each technique has advantages and limitations. We compared the performance of the tube (polyethylene glycol-indirect antiglobulin test [PEG-IAT]) and gel test technique for antibody identification. From January to May 1999, we performed antibody screening tests by gel and tube techniques on 10,123 random blood samples submitted to our reference laboratory. Six hundred and twenty-eight (6.2%) reactive samples were tested for antibody specificity by both methods. One hundred and ninety-six were reactive only by gel: 25 anti-D, 33 anti-C, 76 anti-E, 13 anti-c, 5 anti-e, 18 anti-K, 7 anti-Jka, 2 anti-Dia, 3 anti-S, 8 combination Rh antibodies (1 with anti- K), and 6 other antibody specificities. Two samples were reactive only by PEG-IAT: 1 anti-K and 1 anti-Dia. Four hundred and thirty were positive by the two methods: 156 anti-D, 9 anti-C, 68 anti-E, 15 anti-c, 6 anti-e, 61 anti-K, 12 anti-Jka, 17 anti-Dia, 5 anti-S, 73 combination Rh antibodies (2 with anti-K), and 8 other antibody specificities. Based on this study, the gel test is more sensitive (p <.01) than the tube test for identifying potentially clinically significant antibodies.
    Immunohematology / American Red Cross 02/2000; 16(4):138-41.
Show more