A comparison between human magnetostimulation thresholds in whole-body and head/neck gradient coils

Department of Physics and Astronomy, University of Western Ontario, Ontario, Canada.
Magnetic Resonance in Medicine (Impact Factor: 3.4). 08/2001; 46(2):386-94. DOI: 10.1002/mrm.1202
Source: PubMed

ABSTRACT Gradient coil magnetostimulation thresholds were measured in a group of 20 volunteers in both a whole-body gradient coil and a head/neck gradient coil. Both coils were operated using both x and y axes simultaneously (xy oblique mode). The waveform applied was a 64-lobe trapezoidal train with 1-ms flat-tops and varying rise times. Thresholds were based on the subjects' perception of stimulation, and painful sensations were not elicited. Thresholds were expressed in terms of the total gradient excursion required to cause stimulation as a function of the duration of the excursion. Thresholds for each subject were fit to a linear model, and values for the threshold curve slope (SR(min)) and vertical axis intercept (DeltaG(min)) were extracted. For the body coil, the mean values were: SR(min) = 62.2 mT/m/ms, DeltaG(min) = 44.4 mT/m. For the head/neck coil, the mean values were: SR(min) = 87.3 mT/m/ms, DeltaG(min) = 78.9 mT/m. These curve parameters were combined with calculated values for the induced electric field as a function of position within the coil to yield the tissue specific parameters E(r) (electric field rheobase) and tau(c) (chronaxie). For tissue stimulated within the body coil, the mean values were: E(r) = 1.8 V/m, tau(c) = 770 micros. For tissue stimulated within the head/neck coil, the mean values were: E(r) = 1.3 V/m, tau(c) = 1100 micros. Scalar potential contributions were not included in the calculation of induced electric fields. The mean threshold curves were combined with the gradient system performance curves to produce operational limit curves. The operational limit curves for the head/neck coil system were verified to be higher than those of the whole-body coil; however, the head/neck system was also found to be physiologically limited over a greater range of its operation than was the body coil. Subject thresholds between the two coils were not well correlated.

0 Followers
 · 
83 Views
  • Source
  • [Show abstract] [Hide abstract]
    ABSTRACT: In this study, a boundary element method has been implemented to design and analyze the performance of curved gradient coil geometries as a function of the degree of curvature over all three axes, with designs varying continuously from planar to full cylindrical. It was found that there is a monotonic increase in gradient performance with degree of curvature with little gain beyond half-cylindrical coil geometries for a specific region of interest located 10 cm above the coil surface. The efficiencies of the half-cylindrical geometry coils are 0.76 mT m−1 A−1, 0.71 mT m−1 A−1, and 0.76 mT m−1 A−1 for the x-, y-, and z-gradient axes, respectively, when scaled to 800 μH inductance. The gradient coils presented in this study would serve as anatomically specific gradient channels to be used in conjunction with larger, whole-body coils to comprise a 6-channel hybrid system. The function of these channels could include the ability to provide very high performance diffusion tensor imaging in a specified volume of tissue such as the breast, prostate, or posterior regions of the brain. © 2012 Wiley Periodicals, Inc. Concepts Magn Reson Part B (Magn Reson Engineering) 41B: 62–71, 2012
    Concepts in Magnetic Resonance Part B Magnetic Resonance Engineering 04/2012; 41B(2). DOI:10.1002/cmr.b.21212 · 0.95 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Heterogeneous model of the human body and scalar potential finite difference method are used to compute electric filed in tissue. This field is compared to the previously obtained thresholds for stimulation of peripheral nerves.
    01/2003; 4. DOI:10.1109/IEMBS.2003.1280981

Preview

Download
1 Download
Available from